Post-Market Clinical Investigation of the Clareon® IOL

Not Applicable
Completed
Conditions
Cataract
Aphakia
Interventions
Device: Clareon® IOL
Procedure: Cataract Surgery
Registration Number
NCT03316885
Lead Sponsor
Alcon Research
Brief Summary

The primary purpose of this study is to evaluate the long-term (3 years) visual acuity and adverse event outcomes for the Clareon® Intraocular Lens (IOL). A comparison to historical safety and performance endpoint (SPE) rates as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014 will be conducted at one year. The secondary purpose of this study is to evaluate the visual acuity outcomes with the Clareon IOL at Years 2 and 3.

Detailed Description

Subjects will attend a total of 12 study visits over a period of approximately 36 months: One preoperative screening visit, two operative visits, and nine post-implantation visits. Both eyes will be implanted, with the second eye implantation to occur 2-21 days after the first. Primary endpoint data will be collected at the Year 1 visit (Day 330-420 post-implantation from second eye surgery).

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
245
Inclusion Criteria
  • Diagnosed with bilateral cataracts
  • Planned small incision cataract removal surgery
  • Able to comprehend and willing to sign a statement of informed consent and complete all required post-implantation visits

Key

Exclusion Criteria
  • Subjects who may reasonably be expected to require ocular surgical treatment or refractive surgical procedures at any time during the study
  • Certain eye conditions, as specified in the protocol, including but not limited to glaucoma, diabetic retinopathy, diabetic macular edema, and macular degeneration
  • Pregnant or lactating, current or planned during the course of the study

Other protocol-defined inclusion/exclusion criteria may apply.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Clareon IOLClareon® IOLClareon® aspheric hydrophobic acrylic intraocular lens implanted with the Alcon Monarch III-D delivery system as a replacement of the human crystalline lens during cataract surgery
Clareon IOLCataract SurgeryClareon® aspheric hydrophobic acrylic intraocular lens implanted with the Alcon Monarch III-D delivery system as a replacement of the human crystalline lens during cataract surgery
Primary Outcome Measures
NameTimeMethod
Percentage of First Implanted Eyes Achieving Best-Corrected Distance Visual Acuity (BCDVA) of 0.3 Logarithm Minimum Angle of Resolution (logMAR) or Better at 1 Year Post-ImplantationYear 1 post-implantation from second eye surgery

Visual acuity (VA) was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic Early Treatment Diabetic Retinopathy Study (ETDRS) charts at a distance of 4 meters (m) from the eye. VA was measured in logarithm of the minimum angle of resolution (logMAR), with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical safety and performance endpoint (SPE) of 92.5% \[as reported in the European Standard International Organization for Standardization (EN ISO) 11979-7:2014\].

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 1 Year Post-ImplantationYear 1 post-implantation from second eye surgery

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. VA was compared to the historical SPE of 92.5% (EN ISO 11979-7:2014).

Secondary Outcome Measures
NameTimeMethod
UCDVA at 2 Years Post-Implantation - First Implanted EyeYear 2 post-implantation from second eye surgery

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-ImplantationYear 2 post-implantation from second eye surgery

VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 2 Years Post-ImplantationYear 2 post-implantation from second eye surgery

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Percentage of First Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-ImplantationYear 3 post-implantation from second eye surgery

VA was tested for the first implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

Uncorrected Distance Visual Acuity (UCDVA) at 1 Year Post-Implantation - First Implanted EyeYear 1 post-implantation from second eye surgery

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 diopter (D) spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

UCDVA at 2 Years Post-Implantation - Second Implanted EyeYear 2 post-implantation from second eye surgery

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Percentage of Second Implanted Eyes Achieving BCDVA of 0.3 logMAR or Better at 3 Years Post-ImplantationYear 3 post-implantation from second eye surgery

VA was tested for the second implanted eye using the correction obtained from the manifest refraction and 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A BCDVA of 0.3 logMAR is equivalent to 20/40 Snellen visual acuity. No hypothesis testing was prespecified for this endpoint.

UCDVA at 1 Year Post-Implantation - Second Implanted EyeYear 1 post-implantation from second eye surgery

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

UCDVA at 3 Years Post-Implantation - First Implanted EyeYear 3 post-implantation from second eye surgery

VA was tested for the first implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

UCDVA at 3 Years Post-Implantation - Second Implanted EyeYear 3 post-implantation from second eye surgery

VA was tested for the second implanted eye without visual correction using 100% contrast electronic ETDRS charts at a distance of 4 m from the eye. +0.25 D spherical power was applied to correct for optical infinity. VA was measured in logMAR, with 0.02 logMAR increment corresponding to a single letter on an ETDRS chart. A UCDVA of 0.0 logMAR is equivalent to 20/20 Snellen visual acuity, with a lower numeric value representing better visual acuity. No hypothesis testing was prespecified for this endpoint.

Trial Locations

Locations (1)

Alcon Investigative Site

🇬🇧

London, United Kingdom