Stress Echo and Electric Impedance Tomography (EIT) Pilot Study to Investigate the Influence of Cardiac Output on Parameters of the EIT
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 8
- 试验地点
- 1
- 主要终点
- influence of cardiac output on EIT measurements (feasibility)
研究概览
简要总结
Electric Impedance Tomography (EIT) is a technique based on the injection of small currents and voltage measurements using electrodes on the skin surface generating cross-sectional images representing impedance change in a slice of the thorax.It is a radiation free, non- invasive and portable lung imaging technique. Impedance changes in lung ventilation as well as in lung perfusion can be shown. To investigate lung perfusion by EIT measurements, cardiac output measured by routinely undergone stress echocardiography at rest and stress are compared to parameters measured by EIT.
详细描述
In 15 patients who routinely have to undergo stress echocardiography Electric Impedance Tomography (EIT) measurements are performed parallel to measurements of cardiac output at rest, low stress generated by a dosage of 10µg dobutamine, at peak generated by a dosage between 20-40µg dobutamine (individually)and at relaxation(pulse <100). Therefore prior to the examination a rubber belt with 16 integrated electrocardiographic electrodes is placed around the thorax at the fifth or sixth intercostal space, connected with an EIT device. EIT data are generated by application of a small alternating electrical current of 5mA and 50kHz and are stored and analyzed offline on a personal computer. Changes due to ventilation as well as changes due to perfusion will be measured by EIT. Cardiac output will be retrospectively compared to parameters of the EIT. EIT data are stored in a database and pseudonymized evaluated in relationship to routinely measured stress echocardiography data.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Basic Science
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- Male
- 接受健康志愿者
- 否
入选标准
- •male patients >18 years who routinely have to undergo stress echocardiography
- •ejection fraction >50%
- •able to give informed consent
排除标准
- •restrictive lung diseases
- •active implants and heart pace makers and implanted converter defibrillator
- •foreign metallic objects in the thorax
- •Not adequate quality of imaging of stress echocardiography
研究组 & 干预措施
Stress-Echo and EIT
Patients who routinely undergo stress-echocardiography will additionally be measured by EIT using a rubber belt, which will be placed around their chest
干预措施: EIT (Electrical Impedance Tomography) (Device)
结局指标
主要结局
influence of cardiac output on EIT measurements (feasibility)
时间窗: single measurement of 45 minutes
It will be investigated if a change in cardiac output, which is measured by stress-echocardiography by dobutamine, has an influence on the parameters determined by EIT measurements.
次要结局
未报告次要终点