MedPath

Clinical Validation of mμSORS for Noninvasive Blood Glucose Detection

Not Applicable
Completed
Conditions
Diabetes Mellitus
Medical Device
Interventions
Device: Multi-channel Microspatial Offset Raman Scattering Spectroscopy (mμSORS) for Noninvasive Blood Glucose Detection
Registration Number
NCT05504005
Lead Sponsor
Shanghai Jiao Tong University School of Medicine
Brief Summary

This is a single-center, open-label, prospective study. Blood glucose was measured at different time points during oral glucose tolerance testing in healthy subjects and patients with type 2 diabetes mellitus, using both venous plasma and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS). Venous plasma glucose was set as gold standard.The two measurements were collected synchronously so as to calculate the mean absolute relative difference (MARD) and the consensus error grid (CEG). Accuracy of non-invasive blood glucose testing by mμSORS will be validated. MARD for two measurement methods in different blood glucose ranges and the effect of mμSORS on safety in the patient such as adverse events will also be assessed.

Detailed Description

In order to clinically verify the accuracy and safety of non-invasive blood glucose measurement by the mμSORS technology, and to construct a model, this study will conduct oral glucose tolerance tests on healthy subjects and patients with type 2 diabetes at different time points during the OGTT. There will be 12 points as follows: 0-min and post glucose-load 10-min,30-min,60-min,90-min,120-min,150-min,180-min,210-min,240-min,270-min, and 300-min.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
35
Inclusion Criteria

Not provided

Exclusion Criteria

Not provided

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Clinical Validation of mμSORS for Noninvasive Blood Glucose DetectionMulti-channel Microspatial Offset Raman Scattering Spectroscopy (mμSORS) for Noninvasive Blood Glucose DetectionEnrolled subjects will perform oral glucose tolerance test. A measurement session of blood glucose consists of plasma sample and a measurement by mμSORS will be conducted synchronously.
Primary Outcome Measures
NameTimeMethod
Mean absolute relative deviation (MARD) of venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT.2 months

Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.

Consensus Error Grid (CEG) for venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT.2 months

Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.

Secondary Outcome Measures
NameTimeMethod
Incidence of Treatment-Emergent Adverse Events2 months

Safety in the patients.

MARD for two measurement methods in different blood glucose ranges.2 months

Plasma glucose ranges 3.9-10.0 mmol/L, \<3.9 mmol/L and \>10.0 mmol/L.

Trial Locations

Locations (1)

Ruijin Hospital

🇨🇳

Shanghai, Shanghai, China

© Copyright 2025. All Rights Reserved by MedPath