The Equivalency of the Imp SFB7+ (L-Dex U400) With the Imp XCA Device

ImpediMed Limited1 个研究点 分布在 1 个国家目标入组 30 人开始时间: 2007年11月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
30
试验地点
1
主要终点
Substantial equivalence of the SFB7+ (L-Dex U400) to XCA in the assessment Lymphoedema

研究概览

简要总结

The purpose of this project is to prove that the Imp™SFB7+ (L-Dex U400) is equivalent to the Imp™ XCA device in assessing Lymphoedema. The Imp™ XCA device is a simple device capable of assessing Lymphoedema of a single arm. The Imp™SFB7+ (L-Dex U400) is a more complex device capable of assessing Lymphoedema in a larger population of Lymphoedema sufferers. The Imp™SFB7+ (L-Dex U400) will be developed to assess patients with Lymphoedema in their arms or their legs. The Imp™ XCA and Imp™ SFB7 are both TGA approved devices. The Imp™ SFB7+ (L-Dex U400) is similar to the Imp™ SFB7 the changes are to the software and user interface.

详细描述

When Lymphoedema is present, lymph and other fluids build up in the interstitial spaces of the tissues. This results in an overall increase in the total amount of extracellular fluid (ECF) in the limb, causing swelling. This can be documented by measuring the impedance (opposition) to a low frequency current that has been passed into the limb. Low frequency current travels predominantly through the ECF, where the Lymphoedema manifests. As the fluid builds up in the limb, the impedance to the current decreases and it is in this way that low frequency bioimpedance is able to assess Lymphoedema.

Multi-frequency bio-impedance analysis otherwise known as Bioimpedance Spectroscopy (BIS) has been reported to be effective for the measurement of ECF and sub clinical changes in ECF to predict the onset of Lymphoedema in the arms in studies conducted by Cornish et al. It has been reported by Warren et al that BIS can be used as a reliable and accurate tool for documenting presence of lymphoedema in patients with wither upper- or lower-extremity swelling.

The Imp™ XCA device uses an "impedance ratio" methodology to assess unilateral Lymphoedema of the arm. By this method the unaffected arm acts as an internal and subject specific control. The Imp SFB7+ (L-Dex U400) uses the same principles but at a lower frequency. This study is designed to show the equivalence of the 2 devices in assessing unilateral Lymphoedema of the arm.

研究设计

研究类型
Observational
观察模型
Case Control
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 75 Years(Adult, Older Adult)
性别
Female
接受健康志愿者

入选标准

  • Groups 1 and 2
  • Be female between the ages of 18-75 years.
  • Self-describe general health as satisfactory.
  • Understand the proposed study and be willing and fully able to comply with the study procedures.
  • Be a willing participant and be capable of giving and has given informed written consent for entry into the study.
  • Have been previously diagnosed, by current clinical practice, as having unilateral Lymphoedema of the arm of any severity.
  • Have had a mastectomy or breast conservation surgery with axillary sampling or dissection, with or without adjuvant therapy.

排除标准

  • Have a known heart condition or an implantable device such as a pacemaker or ICD.
  • Have a metallic surgical implant (e.g. total hip replacement) not including small implants such as sternal wires or surgical staples.
  • Suffer from a renal disorder.
  • Be taking diuretic medications.
  • Have consumed large amounts of alcohol or caffeinated beverages within six hours of the study.
  • Have undertaken excessive exercise within two hours of BIA.
  • Have a reported fever of > 38oC at time of screening.
  • Be currently in the fourth week of the menstrual cycle.
  • Be pregnant or currently breastfeeding.
  • Be a relative of any member of study staff or is an employee or a relative of an employee of ImpediMed Ltd.

结局指标

主要结局

Substantial equivalence of the SFB7+ (L-Dex U400) to XCA in the assessment Lymphoedema

时间窗: Single point measure

Observation of difference between L-Dex values of lymphoedema population compared to healthy individuals

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor