MedPath

Beat to Beat A Study in Paediatric, Adolescent and Young Adult Patients Who Are Undergoing or Have Undergone Cancer Therapy to Evaluate the Agreement Between QTc Measured Using 12 Lead Electrocardiogram (ECG) and a Wearable Device ECG

Not Applicable
Completed
Conditions
Neoplasms
Cardiac Arrhythmia
Interventions
Device: ECG application
Device: 12 lead ECG
Registration Number
NCT05615376
Lead Sponsor
Murdoch Childrens Research Institute
Brief Summary

A prospective study in paediatric, adolescent and young adult patients aged 7 to 18 years to evaluate the agreement between QTc measured using 12 lead electrocardiogram (ECG) and the wearable device ECG.

Detailed Description

Rationale: Leveraging existing wearable technology has the ability to scale routine screening strategies across large populations and can facilitate ambulatory care. In recent years, wearable technology has rapidly diffused into consumer markets and provides unique opportunities to engage individuals on health issues of personal interest, and to collect personal health data. One example of such technology is the ability to perform electrocardiogram (ECG) type traces on wearable devices. However, utilization of this data source to improve patient outcomes has not been fully realized in the health care field. An app-based heart ECG algorithm that can be remotely analysed for QTc prolongation with potential to then notify the user of such potential irregularities could lead to widespread screening, detection and initiation of treatment solutions (i.e. correction of electrolytes, alteration of exacerbating medications, cardiac monitoring etc).

Trial Design: This will be a prospective, single arm, experimental, non-significant risk study conducted with the assistance of eligible participants. Parents or mature minors will download the Beat to Beat app, once consented.

Once consented the wearable device will be paired to the parent/guardians phone following the steps provided by the wearable device app. Once the device is paired and the study coordinator has assisted with the download of 'Beat to Beat' app, the patients will then be able to commence the ECG readings. A 12 lead ECG will be performed by the study team and immediately following this, an ECG will be recorded on the wearable device. The wearable device will be placed on the left wrist for a Lead I reading, on the left lower abdomen for a Lead II reading and in the fourth intercostal space left parasternal for a chest V2 Lead reading. Recording of the 12 lead ECG and wearable device ECG will be repeated again 4 days later. Once both timepoints (and two device recordings for each time period) have been completed the ECGs will be retrieved from the Beat to Beat app. Both the 12 lead and wearable device ECG will be de-identified and given a patient identifier before being uploaded into the REDCap database.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
45
Inclusion Criteria
  • Paediatric or adolescent diagnosis of cancer AND an inpatient of The Royal Children's Hospital, Melbourne for at least 5 days at the time of consent.

  • Patient age ≥ 7 years at time of eligibility screening

  • If age < 18 years, parent or guardian able to provide consent

  • Parental or participant ( if > 18 years of age) possession of the following at time of eligibility screening, ascertained from automatic hardware/software/device pairing check:

    1. iPhone (6s or later) with iOS version 15.0 or later defined as iPhone model/iOS version used to complete screening eligibility.
    2. Proficient in written and spoken English, defined by self-report of comfort reading, writing, and speaking English on iPhone.
    3. Valid phone number associated with iPhone, ascertained from self-report.
    4. Valid email address, ascertained from self-report.
    5. Participants or parents need to be able to press down on the wearable device crown for 30 seconds
Exclusion Criteria
  • Unable to wear the wearable device.
  • < 18 years of age without guardian or parent to provide consent.
  • Interpreter required for consent purposes.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
ECG App armECG applicationParticipants will be provided with a wearable device on Day 1 and Day 4 whilst admitted in hospital.
ECG App arm12 lead ECGParticipants will be provided with a wearable device on Day 1 and Day 4 whilst admitted in hospital.
Primary Outcome Measures
NameTimeMethod
Recording of a 12 lead ECG and wearable device ECGDay 4 of inpatient stay

12 lead ECG and wearable device ECG recordings will be taken at the time point of care. The wearable device ECG will be placed on the left wrist to record a V1 (lead I) ECG and on the left ankle to record a V2 (lead II) ECG reading from which the QT interval can be calculated.

Calculation of QT interval by two blinded health professionalsDay 4 of inpatient stay

The wearable device ECG and 12 lead ECG will be de-identified and given appropriate study numbers. This data will be listed separately for each patient. The QT interval for each recording will be calculated. This is the time corresponding to beginning of depolarization to repolarization of the ventricles. It is calculated using a standardised approach (Fridericia's formula:QTcF = QT divided by cube root of RR).

Secondary Outcome Measures
NameTimeMethod
To calculate the interobserver variability between the two health care professional readings of QTc.Day 4 inpatient stay

The corrected QT interval will be calculated on both the wearable device ECG and the 12 lead ECG for each patient and measured by a health professional. QT interval will be calculated by the health professional by using the QT and RR interval from the ECGs using Frederichia's interval (QTcF = QT divided by cube root of RR). This data will be collected separately for each patient for each patient The standard deviation of the difference in QTc interval between the two observers will be reported.

Sensitivity calculations of wearable device vs 12 Lead ECGDay 4 inpatient stay

Sensitivity is calculated only in the participants who are declared to have QTc prolongation (abnormal QTc) and is calculated as the proportion patients with QTc prolongation on the wearable device among those with QTc prolongation on the 12-lead ECG.

Specificity calculations of wearable device vs 12 Lead ECGDay 4 inpatient stay

Specificity is calculated only in participants who do not have QTc prolongation and is calculated as the proportion patients who do not have QTc prolongation on the wearable device among those who do not have QTc prolongation on the 12-lead ECG.

Number of Participants with Abnormal QTc that is greater than 0.48mm on 12 lead ECG and wearable device ECGDay 4 inpatient stay

From the analysis of QT measurements from Outcome 2, prolonged or abnormal QTc measurement will be noted. An abnormal QTc is greater than 0.48mm. This data will be collected separately for each patient.

Trial Locations

Locations (1)

The Royal Children's Hospital

🇦🇺

Parkville, Victoria, Australia

© Copyright 2025. All Rights Reserved by MedPath