Comparison of Woven EndoBridge Device Sizing With Conventional Measurements and Virtual Simulation Using the Sim&Size Software: A Multi-Center Experience

University Hospital, Montpellier1 个研究点 分布在 1 个国家目标入组 186 人开始时间: 2015年1月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
186
试验地点
1
主要终点
Median duration of the intervention

研究概览

简要总结

Selection of the appropriate size of the device is mandatory during aneurysm treatment with a WEB. The Investigators aimed to investigate if virtual simulation with Sim&Size software may have an impact on technical, angiographic, and clinical outcomes after WEB treatment. Data from two large-volume centers were collected and compared (January 2017-January 2020). Virtual simulation was systematically adopted in one center, while conventional sizing was used in the other one. Outcomes were the duration of intervention, the radiation dose (mGy), the number of corrective interventions for inappropriate WEB size, the number of WEBs not deployed, angiographic occlusion, and complications. Univariate and multivariate linear models were adopted.

详细描述

Flow disruption with Woven EndoBridge (WEB) is an innovative treatment that has been evaluated in several prospective and retrospective studies1-3. The efficacy and safety of this strategy has been largely proved, and nowadays, indications for WEBs are progressively enlarged. Appropriate sizing of the device is the key factor for treatment with a WEB. Undersized devices may lead to insufficient coverage of the aneurysm neck and inadequate filling of the volume of the sac, lowering the angiographic success. On the other hand, an excessively oversized WEB may cause an unwarranted protrusion into the parent vessel, increasing the risk of ischemic events. The conventional approach for WEB sizing is based on two-dimensional digital subtraction angiography (2D-DSA), adding 1mm to the average width (to assure good wall apposition), and subtracting 1mm to the average height of the aneurysm (to adjust for the longitudinal increase caused by the horizontal compression)4. However, the operator has to anticipate the behavior of the device, predicting tridimensional modifications of the WEB in relation to the volume and shape of the aneurysm dome and neck. Accordingly, appropriate WEB sizing could be difficult and may require experience. Data about software-based simulation to predict changes after device implantation have been reported for flow diversion and braided stents5-7. The Sim&Size software (Sim&Cure, Grabels, France) is able to perform virtual sizing before WEB deployment using pre or per-operative three-dimensional rotational angiography (3D-RA) acquisition. The virtual simulation helps to select the most appropriate size of the device, predicting the behavior of the WEB inside the aneurysm dome, and the surface of the WEB apposition, anticipating protrusion over the neck or inappropriate coverage of the ostium. This multicenter study aimed to explore if the use of the Sim&Size software during endovascular embolization of intracranial aneurysms with WEB devices was associated with measurables procedural advantages compared with conventional sizing methods.

The primary investigated outcomes were 1) the median duration of the intervention; 2) the median radiation dose; 3) the need of corrective interventions; and 4) the number of WEBs opened but finally not deployed because of the inappropriate size of the device.

The secondary investigated endpoints were 1) the overall rate of complications (peri-procedural and long-term adverse events); 2) the morbidity rate (complications associated with permanent change of the clinical and neurological status of the patient); and 3) short-term (6 months) and long-term (12 months or more) aneurysm occlusion and WEB-shape change ("compaction") (decrease in height of the device or a deepening of the proximal and distal concave recesses during follow-up9 based on non-subtracted images)

研究设计

研究类型
Observational
观察模型
Case Control
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

Median duration of the intervention

时间窗: 1day

the median duration of the intervention

Median radiation dose

时间窗: 1 day

the median radiation dose

Need of corrective interventions

时间窗: 1 day

the need of corrective interventions

number of WEBs opened

时间窗: 1 day

the number of WEBs opened but finally not deployed because of the inappropriate size of the device.

次要结局

  • Overall rate of complications(1 day)
  • morbidity rate(1 day)
  • Rate of Aneurysm occlusion(6 months and 12 months)
  • WEB-shape change(6 months and 12 months)