Epicardial Echocardiography of Coronary Anastomoses Using the Echoclip Device
- Conditions
- Coronary Arteriosclerosis
- Interventions
- Device: Echoclip
- Registration Number
- NCT02919124
- Lead Sponsor
- Aalborg University Hospital
- Brief Summary
The aim of this study is to evaluate whether use of the echoclip device (an ultrasound transducer positioning device which can stabilize the involved part of the myocardium on the beating heart) facilitates imaging of coronary bypass anastomoses during coronary bypass surgery. A total of 100 low risk patients undergoing elective on-pump coronary bypass surgery will be included in the study in order to evaluate if the surgeons can visualize the coronary anastomoses before closure of the sternum. Ultrasonograpic pictures will be analyzed directly peroperatively and electronically post-operatively in order to evaluate if selected areas of the anastomoses can be visualized. Use of the echoclip devise will be considered a success if at least 80% of the anastomoses can be visualized.
- Detailed Description
The technical quality of coronary artery bypass anastomoses during coronary artery bypass grafting is often evaluated by measuring flow rate in grafts using transit time flowmetry. Unfortunately, flow rate and flow waveform are poor indicators of the anatomy of anastomoses. A better way to evaluate the anatomy of an anastomosis is 3-dimensional imaging. Three problems must be overcome in performing epicardial echocardiography on the beating heart:
1. The transducer must be kept in a steady position relative to the anastomosis for 5 to 10 cardiac cycles to minimize frame-rate variability.
2. Ultrasound gel must be kept in contact with the transducer and the region studied.
3. The transducer must not cause any deformation of the vessels.
The investigators designed an ultrasound transducer positioning device, the Echoclip device (Aalborg Hospital, Aalborg, Denmark), which can stabilize the involved part of the myocardium on the beating heart, keep the gel at place, and position the ultrasound transducer correctly for imaging. The present study is a feasibility study with the aim to evaluate if the echoclip device facilitates imaging of coronary bypass anastomoses during coronary artery bypass surgery in humans to the same degree as was shown in animal studies. Ultrasonography will be used for visualizing all coronary anastomoses during 100 elective low-risk (logistic II EuroSCORE \<6) on-pump coronary artery bypass procedures. It will be registered if the heel, the central portion and the toe of the anastomosis can be visualized in end-to-side and side-to-side anastomoses.Use of the echoclip devise will be considered a success if at least 80% of all parts of the anastomoses can be visualized either directly or by electronic reading using a special developed algorithm that may identify the inner border of the anastomoses.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 56
- Elective on-pump coronary artery bypass surgery
- Surgery planned to be performed by specialist in cardiac surgery.
- Logistic euroSCORE II <6%.
- Must be able to read and understand the Danish
- Written informed acceptance to participate
- Planned off-pump surgery.
- Peroperativ conversion to off-pump surgery.
- Logistic EuroSCORE II ≥ 6.
- The patient do not understand written or spoken Danish.
- Surgery within 24 hours after admittance to hospital
- Patients in whom the surgeon wants to end the surgery in a hurry due to e.g. hemodynamic problems or peroperative complications.
- The surgeon is not a specialist in cardiac surgery, e.g. surgeon in training.
- Pregnant or brest feeding patient.
- No written acceptance to participate.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Echoclip device Echoclip Ultrasonography using the echoclip device
- Primary Outcome Measures
Name Time Method Quality of periferal coronary artery bypass anastomotic structures Peroperative Ultrasonographic images of periferal coronary artery bypass anastomoses will be studied in order to evaluate if the area of anastomotic structures from various locations of the anastomoses (heel, central portion and the to toe) can be visualized directly on the screen using the echoclip device.
- Secondary Outcome Measures
Name Time Method Automatic quality control of periferal coronary artery bypass anastomotic structures Peroperative To quantify the anastomotic quality from epicardial utrasound images obtained peroperative, the area of anastomotic structures from various locations of the anastomoses (heel, central portion and the to toe) will be meassured using an automatic anastomosis segmentation algorithm to extract the area of the anastomotic structures. The ultrasound images obtained peroperative will be analyzed in the Laboratory when 50 patients and 100 patients have been included in the study, respectively.
Trial Locations
- Locations (1)
Aalborg University Hospital
🇩🇰Aalborg, Denmark