Respiratory Rate Validation Study - ChroniSense Polso
Overview
- Phase
- Not Applicable
- Intervention
- Not specified
- Conditions
- Respiratory Rate
- Sponsor
- ChroniSense Medical Ltd.
- Enrollment
- 20
- Locations
- 1
- Primary Endpoint
- Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate
- Status
- Completed
- Last Updated
- 4 years ago
Overview
Brief Summary
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso
Detailed Description
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso. A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers. Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.
Investigators
Eligibility Criteria
Inclusion Criteria
- •Subject must have the ability to understand and provide written informed consent
- •Subject is adult over 18 years of age
- •Subject must be willing and able to comply with study procedures and duration
- •Subject is a non-smoker, or a smoker who has refrained from smoking for 1 day.
- •Male or female of any race
Exclusion Criteria
- •Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
- •Subjects with known respiratory conditions such as:
- •uncontrolled / severe asthma,
- •pneumonia / bronchitis,
- •shortness of breath / respiratory distress,
- •respiratory or lung surgery,
- •emphysema, COPD, lung disease
- •Subjects with self-reported heart or cardiovascular conditions such as:
- •high blood pressure: systolic \>140 mmHg or diastolic \>90 mmHg
- •have had cardiovascular surgery
Outcomes
Primary Outcomes
Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate
Time Frame: Through study completion, 1 month average
Accuracy of Polso Monitoring System (DUT) relative to a reference End Tidal Carbon Dioxide monitor-derived respiration rate (REF). Polso reports respiration rate (breaths per minute) directly, at 60 second intervals. End Tidal Carbon Dioxide monitor waveforms will be hand-scored, based on respiration waveform peaks, to derive respiration rate (breaths per minute). The readings from each device will be synchronized to achieve paired simultaneous respiration rates, and aggregated into calculation of a comparative accuracy measure, namely Accuracy root-mean-square (Arms) difference, in breaths per minute, between the Polso (DUT) and the reference (REF) respiratory rate system, for all stable respiratory periods. Acceptance criteria Arms \<3.