NCT05246358
Completed
Not Applicable

Respiratory Rate Validation Study - ChroniSense Polso

ChroniSense Medical Ltd.1 site in 1 country20 target enrollmentMay 10, 2021

Overview

Phase
Not Applicable
Intervention
Not specified
Conditions
Respiratory Rate
Sponsor
ChroniSense Medical Ltd.
Enrollment
20
Locations
1
Primary Endpoint
Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate
Status
Completed
Last Updated
4 years ago

Overview

Brief Summary

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso

Detailed Description

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso. A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers. Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.

Registry
clinicaltrials.gov
Start Date
May 10, 2021
End Date
June 11, 2021
Last Updated
4 years ago
Study Type
Interventional
Study Design
Single Group
Sex
All

Investigators

Responsible Party
Sponsor

Eligibility Criteria

Inclusion Criteria

  • Subject must have the ability to understand and provide written informed consent
  • Subject is adult over 18 years of age
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker, or a smoker who has refrained from smoking for 1 day.
  • Male or female of any race

Exclusion Criteria

  • Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • Subjects with known respiratory conditions such as:
  • uncontrolled / severe asthma,
  • pneumonia / bronchitis,
  • shortness of breath / respiratory distress,
  • respiratory or lung surgery,
  • emphysema, COPD, lung disease
  • Subjects with self-reported heart or cardiovascular conditions such as:
  • high blood pressure: systolic \>140 mmHg or diastolic \>90 mmHg
  • have had cardiovascular surgery

Outcomes

Primary Outcomes

Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate

Time Frame: Through study completion, 1 month average

Accuracy of Polso Monitoring System (DUT) relative to a reference End Tidal Carbon Dioxide monitor-derived respiration rate (REF). Polso reports respiration rate (breaths per minute) directly, at 60 second intervals. End Tidal Carbon Dioxide monitor waveforms will be hand-scored, based on respiration waveform peaks, to derive respiration rate (breaths per minute). The readings from each device will be synchronized to achieve paired simultaneous respiration rates, and aggregated into calculation of a comparative accuracy measure, namely Accuracy root-mean-square (Arms) difference, in breaths per minute, between the Polso (DUT) and the reference (REF) respiratory rate system, for all stable respiratory periods. Acceptance criteria Arms \<3.