Inter- and Intra-Observer Variability in Transcranial Doppler (TCD) Technique in Neurocritical Care Patients
试验速览
- 阶段
- 不适用
- 状态
- 终止
- 入组人数
- 3
- 试验地点
- 1
- 主要终点
- Determine the Inter-variability of TCD Ultrasound - EDV Measurement
研究概览
简要总结
This study aims to determine the inter- and intra-variability of Transcranial Doppler (TCD) ultrasound in neuro-critical care patients who are planned for consecutive daily TCD evaluations.
详细描述
Patients enrolled in the study are recruited from the pool of neuro-critical care patients admitted to the surgical intensive care unit (SICU). When the patients are admitted in the surgical ICU, the study staff will determine eligibility based on the protocol inclusion criteria. Due to the nature of the disease, there is a high likelihood that patients are either unconscious, of limited mental capacity and/or have been intubated therefore when appropriate the legally authorized representative will be utilized to obtain research consent.
To meet study inclusion, transcranial Doppler (TCD) testing will be ordered as part of the standard of care for these patients. The test will not be ordered solely for research purposes. There are no known side effects from the non-invasive measurement of cerebral blood flow using ultrasound
Three different TCD technicians will do triplicate readings on 3 consecutive days on up to 12 patients already undergoing TCD as ordered by their treating team. Standard of care on specific neuro critical care patients (such as cerebral aneurysms) is to undergo daily TCD monitoring to assess for possible vasospasm. Patients will be in the supine position while measurements are obtained. The probe will be placed in the preauricular region of the temporal window
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Adults 18 years of age or older
- •Current hospitalization for a neurologic issue and admitted to the surgical intensive care unit.
- •Undergoing daily transcranial Doppler imaging No
- •English speaking patients and/or legally authorized representative (LAR)
排除标准
- •Less than 18 years of age
- •Patient is unable to obtain consent and no LAR is identified to provide consent
- •Non-English speaking patient or LAR
- •Unable to perform the study due to lack of availability of TCD technologists
结局指标
主要结局
Determine the Inter-variability of TCD Ultrasound - EDV Measurement
时间窗: 3 consecutive days
The inter-variability (agreement among 3 TCD technicians) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated based on the readings from the same visit of a patient by different technician as the index for the interrater reliability or agreement among the three technicians. Based on the suggestion from Koo and Li (2016), the interpretation of the agreement by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
Determine the Inter-variability of TCD Ultrasound - PFV Measurement
时间窗: 3 consecutive days
The inter-variability (agreement among 3 TCD technicians) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated based on the readings from the same visit of a patient by different technician as the index for the interrater reliability or agreement among the three technicians. Based on the suggestion from Koo and Li (2016), the interpretation of the agreement by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
Determine the Inter-variability of TCD Ultrasound - Depth Measurement
时间窗: 3 consecutive days
The inter-variability (agreement among 3 TCD technicians) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated based on the readings from the same visit of a patient by different technician as the index for the interrater reliability or agreement among the three technicians. Based on the suggestion from Koo and Li (2016), the interpretation of the agreement by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
Determine the Intra-variability of TCD Ultrasound - EDV Measurement
时间窗: 3 Consecutive Days
The intra-variability (reliability or consistency within each TCD technician) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated for each technician based on the readings from the same patient. Based on the suggestion from Koo and Li (2016), the interpretation of the reliability by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
Determine the Intra-variability of TCD Ultrasound - PFV Measurement
时间窗: 3 Consecutive Days
The intra-variability (reliability or consistency within each TCD technician) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated for each technician based on the readings from the same patient. Based on the suggestion from Koo and Li (2016), the interpretation of the reliability by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
Determine the Intra-variability of TCD Ultrasound - Depth Measurment
时间窗: 3 Consecutive Days
The intra-variability (reliability or consistency within each TCD technician) was measured by using the intraclass correlation coefficient (ICC). Three technicians took measurements from the same patient at three different visits. The ICC is calculated for each technician based on the readings from the same patient. Based on the suggestion from Koo and Li (2016), the interpretation of the reliability by using the ICC is defined as: below 0.50 = poor, 0.50-0.75 = moderate, 0.76-0.90 = good, above 090 = excellent.
次要结局
- Change in Velocity Prompting Clinical Intervention(3 Consecutive Days)