GlucoClear ICU Study
- Conditions
- Intensive Care Unit Blood Glucose Monitoring
- Interventions
- Device: GlucoClear System
- Registration Number
- NCT02238041
- Lead Sponsor
- Edwards Lifesciences
- Brief Summary
The primary purpose of this Study is evaluation of the safety and accuracy of the GlucoClear CGM System.
- Detailed Description
Design: Non-randomized, non-treatment, prospective, open label Study.
The GlucoClear System will not display real-time glucose values, glucose trend graphs, or glucose alarms to either the Study subjects or the Study site clinical personnel. There will be no treatment of any subject based on the output of the System.
After providing written informed consent, subjects meeting inclusion/exclusion criteria will be monitored by the System for a maximum of 72 hours.
Up to 51 1 mL blood samples will be drawn from each subject for comparative glucose measurements on blood glucose analyzers over the 72 hour period.
After the GlucoClear Sensor removal, subjects will be contacted for a follow-up assessment 14 days later, and any subsequent adverse events will be documented.
Recruitment & Eligibility
- Status
- WITHDRAWN
- Sex
- All
- Target Recruitment
- Not specified
- 18 years Surgical ICU or surgical ICU plus intermediate care unit stay ≥ 24 hours Signed consent
Transplant patient End stage medical conditions or diseases Restricted venous access History of HIT or heparin allergy Contraindication for anti-coagulation Positive pregnancy test In another study
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Arm && Interventions
Group Intervention Description GlucoClear System GlucoClear System -
- Primary Outcome Measures
Name Time Method 15/20% Accuracy During 72 hours of anticipated Sensor dwell time. 15/20% Accuracy: The proportion of GlucoClear glucose measurements within ± 15 mg/dL of Comparator values \< 75 mg/dL and within ± 20% of Comparator values ≥ 75 mg/dL)
- Secondary Outcome Measures
Name Time Method 12/15% Trend Accuracy During 72 hours of anticipated Sensor dwell time. 12/15% Trend Accuracy: The proportion of GlucoClear glu-cose measurements within ± 12 mg/dL of the interpolated value between two sequential Com-parator values if that interpolat-ed comparator value is \< 80 mg/dL and within ± 15% of in-terpolated Comparator values ≥ 80 mg/dL).