临床试验/NCT00746356
NCT00746356已完成不适用

A Clinical Evaluation of the Current™ Radio Frequency (RF) Implantable Cardioverter Defibrillator (ICD) Models 2215/1215 and Promote™ RF Cardiac Resynchronization Therapy Defibrillator (CRT-D) Model 3215 Pulse Generators

Abbott Medical Devices10 个研究点 分布在 1 个国家目标入组 128 人开始时间: 2008年8月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
128
试验地点
10
主要终点
Right Ventricular (RV) AutoCapture Effectiveness Endpoint - Difference Between the Automatic and Manual Capture Threshold Test

研究概览

简要总结

This study is a prospective, non-randomized, multi-center clinical trial that will assess the safety and effectiveness of investigational models of the Promote™ RF CRT-D and Current™ RF ICD devices. These new ICD and CRT-D devices have new features that allow the device to periodically check how much energy is needed to pace a patient's heart and to automatically adjust the amount of energy used to pace the heart.

详细描述

All patients enrolled in the study will have their device tested to determine if the device can automatically measure how much energy is needed to pace the patient's heart. If the test is successful, the patient's device will be programmed to turn that feature on. 3 months after enrollment, the patient will return to the clinic to have their device re-tested using an automatic test method to measure the amount of energy needed to make the patient's heart beat. A manual test method will also be used to measure the amount of energy needed to make the patient's heart beat. The intent of the study is to show the automatic method works just as well as the manual method.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Parallel
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Have an approved indication, as outlined by American Heart Association/Heart Rhythm Society (AHA/HRS) guidelines, for implantation of an ICD or CRT-D for the treatment of life-threatening ventricular tachyarrhythmia(s) or heart failure, or undergo revision of their ICD or CRT-D system to replace the pulse generator.
  • Have the ability to provide informed consent for study participation and be willing and able to comply with the prescribed follow-up tests and schedule of evaluations.

排除标准

  • Have a classification of Status 1 for cardiac transplantation or consideration for transplantation over the next 3 months.
  • Have had a recent myocardial infarction, unstable angina or cardiac revascularization (percutaneous transluminal coronary angioplasty (PTCA), Stent or Coronary Artery Bypass Graft(CABG)) within 40 days of enrollment.
  • Have had a recent cerebrovascular accident (CVA) or transient ischemic attack (TIA) within 3 months of enrollment.
  • Are allergic to dexamethasone sodium phosphate (DSP).
  • Be currently participating in a clinical investigation that includes an active treatment arm.
  • Be pregnant or are planning for pregnancy within 6 months following enrollment.
  • Have a life expectancy of less than 6 months.
  • Be less than 18 years of age.

研究组 & 干预措施

Promote RF CRT-D

Experimental

Patients with CRT-D device will have the autocapture features of the device tested.

干预措施: Testing of the autocapture features of the device (Device)

Current RF ICD

Experimental

Patients with ICD device will have the autocapture features of the device tested.

干预措施: Testing of the autocapture features of the device (Device)

结局指标

主要结局

Right Ventricular (RV) AutoCapture Effectiveness Endpoint - Difference Between the Automatic and Manual Capture Threshold Test

时间窗: 3 months post implant

Right Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.

Left Ventricular (LV) AutoCapture Effectiveness Endpoint - Difference Between the Automatic and Manual Capture Threshold Test

时间窗: 3 months post implant

Left Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.

Atrial AutoCapture (ACap) Confirm Effectiveness Endpoint - Absolute Difference Between the Automatic and Manual Capture Threshold Test

时间窗: 3 months post implant

Atrial AutoCapture Confirm is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the atrial pacing voltage until it determines that the device is no longer capturing the atria. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the atrial pulse voltage and determines the atrial threshold by viewing the EKG.

Ventricular Autocapture Effectiveness Endpoint - Absolute Difference Between the Automatic and Manual Capture Threshold Test

时间窗: 3 months post implant

Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.

Percentage of Participants Free of System-related Complications at 3-months Post Implant

时间窗: 3 months post implant

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor