临床试验/NCT02433210
NCT02433210已完成不适用

A Clinical Study Comparing the Basic Performance and Blood Compatibility Characteristics of Nipro ELISIO-H, Gambro Polyflux Revaclear and Fresenius Optiflux Dialyzers

University of Missouri-Columbia2 个研究点 分布在 1 个国家目标入组 11 人开始时间: 2015年6月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
11
试验地点
2
主要终点
Hemocompatibility as Shown by C5a for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers of Dialyzers

研究概览

简要总结

The aim of the study is to confirm the safety and clinical effectiveness of Nipro ELISIO-15H and to compare its performance characteristics and hemocompatibility with those of the commercially available dialyzers in U.S.A., Gambro Polyflux Revaclear and Fresenius F160NRe.

10 stable patients will be dialyzed 3 times/week with each dialyzer over 3 weeks. Blood samples for several parameters will be collected during hemodialysis to assess the performance and hemocompatibility of the dialyzers under study. Data thus collected will be analyzed for comparison by standard statistical analysis.

详细描述

Current patients receiving dialysis treatment for end stage renal disease and who meet the inclusion/exclusion criteria will be entered into the study. These patients will continue on their standard dialysis regimen as stated in the clinics Standard Operating Procedures. Each patient will be studied for three weeks with each week starting with a dialyzer of a different manufacture but of comparable characteristics. Solute clearances and removal will be determined at each of the three dialysis sessions during the week. During session 2 of each week additional specimens will be taken to analyze for hemocompatibility. At the end of the study all clearances and the hemocompatibility markers for each dialyzer will be compared to one another as well as to the manufactures stated characteristics. Blood for the clearance and removal calculations will be taken at time 0, 60 minutes and 240 minutes. Blood for the hemocompatibility determinations will be taken at time 0, 15, 30, 60 and 240 minutes. Flow rates, anticoagulant therapy, erythropoietin therapy, venous and arterial pressures will also be recorded.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Basic Science
盲法
None

盲法说明

Open Label

入排标准

年龄范围
18 Years 至 85 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • ESRD patients 18 years or older
  • Stable on hemodialysis for more than 3 months
  • Stable hemoglobin between 11-12 g/dl
  • Stable arteriovenous (AV) fistula vascular access
  • Stable anticoagulation and ESA regimen
  • No active infection
  • Able to sign informed consent and able to participate in the study
  • Medically stable

排除标准

  • Participation in another study which may interfere with the planned study
  • Active infection
  • Medical conditions which may interfere with the study (cardiac, liver disease, Hepatitis)
  • Allergy to dialyzer membrane materials e.g. polysulfone
  • Patients which cannot tolerate Heparin
  • Female who are pregnant or planning to be pregnant
  • Problem with or allergy to anticoagulation

结局指标

主要结局

Hemocompatibility as Shown by C5a for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers of Dialyzers

时间窗: Once a week during the second of the three weekly sessions specimens will be taken at time points 0, 15, 30, 60 and 240 minutes. Measurements, all patients, were summed for each dialyzer type by time period.

The molecules monitored will be C5a, thrombin/anti-thrombin complex and a complete blood count with platelets.

Hemocompatibility as Shown by Platelets for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers

时间窗: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time periods. All measurements for all patients were summed for each dialyzer type and time point.

Measure of the % change from Baseline for the hemocompatibility marker Platelets for the Optiflux, Revaclear, and ELISIO dialyzers.

Hemocompatibility as Shown by Hematocrit for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers

时间窗: Once a week during the second of three weekly sessions samples will be taken at the 0, 15, 30, 60, and 240 minute timepoints. Measurements for all patients were summed for each dialyzer type by time period for analysis.

Measure of the % change from Baseline for the hemocompatibility marker Hematocrit for the Optiflux, Revaclear, and ELISIO dialyzers.

Hemocompatibility as Shown by White Cell Count for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers

时间窗: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time periods. All measurements for all patients were summed for each dialyzer type and time point.

Measure of the % change from Baseline for the hemocompatibility marker White Cell Count for the Optiflux, Revaclear, and ELISIO dialyzers.

Clearance and Solute Removal Characteristics of Dialyzers

时间窗: During each of the three weekly dialysis sessions measurements of urea were made at the 60 minute time point for each patient in the study. At the end of the study all measurements were summed for each of the three dialyzers.

Clearances of small, middle and large molecular weight solutes will be measured at 60 minute time point utilizing urea, creatinine, beta 2 microglobulin, phosphate and myoglobin. Dialyzer clearance is calculated as follows: 60 minute clearance:(((\[S\]60'Art - \[S\]60'Ven) / \[S\]60'Art) \* Qb) where \[S\] = solute concentration; Qb = blood flow rate. S

Hemocompatibility as Shown by Thrombin/Antithrombin Complex for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers.

时间窗: Once a week during the second of the three weekly sessions specimens will be taken at the 60 and 240 minutes timepoints. Measurements for all patients were summed for each dialyzer type by time period for analysis..

Measure of the % change from Baseline for the hemocompatibility marker Thrombin/Antithrombin complex for the Optiflux, Revaclear, and ELISIO dialyzers.

Hemocompatibility as Shown by Hemoglobin for Optiflux, Revaclear, and ELISIO Hemodialysis Dialyzers

时间窗: Once a week during the second of the three weekly sessions samples will be taken at the 0, 15, 30, 60 and 240 minutes time points. Measurements for all patients were summed for each dialyzer type and time point for analysis.

Measure of the % change from Baseline for the hemocompatibility marker Hemoglobin for the Optiflux, Revaclear, and ELISIO dialyzers.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Madhukar Misra

Professor of Clinical medicine

University of Missouri-Columbia