临床试验/NCT02644837
NCT02644837终止不适用

Randomised Crossover Comparison of the Ambu AuraGain and the iGel in Anaesthetised Adults

NHS Tayside2 个研究点 分布在 1 个国家目标入组 2 人开始时间: 2016年10月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
终止
发起方
NHS Tayside
入组人数
2
试验地点
2
主要终点
oropharyngeal leak pressure

研究概览

简要总结

This study will look at key performance indicators of 2 supraglottic airway devices in anaesthetised adults in a crossover manner.

详细描述

A supraglottic airway device (SAD), also referred to as laryngeal mask airway (LMA), is a medical device that maintains a patients airway, allowing unobstructed ventilation during anaesthesia. They are designed to sit in the patient's hypopharynx (throat), with their elliptical head forming a seal around the supraglottic structures (the patients "voice box"). They have been in common use since 1989, transforming anaesthetic practice and are now the predominant airway device within anaesthesia, being used in approximately 56% of all general anaesthetic cases in the UK(1). Their popularity for routine use stems from its perceived benefits over traditional forms of airway management coupled with their high overall success rate and low complication rate.

SADs, in anaesthetic practice, are inserted after induction of general anaesthesia, by or under supervision of, a trained medical professional. All patients are fully monitored as set out by the Association of Anaesthetist of Great Britain and Ireland; Recommendations for standards of monitoring during anaesthesia and recovery (2). Devices are used in concordance with manufacturer instruction. Successful first time placement is achieved in a large majority of patients and allows provision of oxygen and ventilation. In some patients, one device or size of device may present sub-optimal performance resulting in the removal of the device. Further actions in this scenario would include attempting a second insertion of the same device, attempting an insertion with an alternative size of the same device or a further attempt with an alterative design of device in order to achieve a patent airway. Further attempts may be taken but should be limited to avoid unnecessary trauma. In some cases the use of the SAD is abandoned and the patient is intubated with an endotracheal tube in order to provide a safe secure airway.

Alternatives to use of a SAD for airway management include use of a simple facemask or tracheal intubation with an endotracheal tube. Advantages of using a SAD for anaesthetic airway management compared to a facemask are improved oxygen saturation and less operator/anaesthetist fatigue. Compared to an endotracheal tube the advantages of using a SAD are improved haemodynamic stability at induction and emergence from anaesthesia, reduced anaesthetic requirements, improved quality of emergence, lower incidence of sore throat, increased ease of placement and reduced risk of dental damage. Risks of using an LMA compared to an endotracheal tube are increased gastric insufflation and aspiration with these being more common in patients who are poorly selected for anaesthesia with a SAD. Sore throat is the most common complication with a study

There have been significant developments in design in the SAD market in recent years with development of a "second generation" range of devices involving new materials and designs that integrate protective bite blocks, gastric drainage channels and improved supraglottic seal enhancing patient safety when using these devices. There are now some new "third generation" SADs on the market with designs that may enhance anaesthetic practice and improve patient safety further.

This study will follow similar ethically approved protocols used in published studies comparing airway equipment (3,4). These are only 2 examples but there are hundreds of LMA studies throughout the literature. The Ambu AuraGain is a new device and has not been studied fully. Comparative studies frequently assess oropharyngeal leak pressure (OLP), a fibre optic assessment of glottic alignment, ease of insertion, ability to insert a nasogastric tube down the gastric port and frequency of manipulations.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • patients who have read and understood patient information leaflets about the study and undergo informed consent.
  • patients whose anaesthesia management is being carried out by one of the named investigators
  • adult (age >18 years) patients undergoing general anaesthesia
  • American Society of Anaesthesiology Grading (ASA) 1-3
  • suitability for general anaesthesia using a laryngeal mask airway device
  • Exclusion Criteria;
  • the presence of significant acute or chronic lung disease
  • pathology of neck or upper respiratory tract
  • an identified or anticipated difficult intubation
  • an increased risk of aspiration (hiatus hernia, gastro-oesophageal reflux or full stomach etc.) pregnant women
  • a body mass index greater than 35kg.m-2
  • patients unable to communicate fully in English.

排除标准

  • 未提供

研究组 & 干预措施

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Insertion of Ambu AuraGain laryngeal mask airway and Igel (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Fibreoptic assessment with Ambu A-scope (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Ability to insert nasogastric tube (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Measurement of OLP (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Ability to perform positive pressure ventilation (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Assessment of ease of insertion (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Record number of manipulations (Device)

i-Gel and AuraGain

Active Comparator

In this arm of the crossover study, patients will undergo insertion of Ambu AuraGain laryngeal mask airway and iGel and will undergo initial management with the iGel. Primary and Secondary outcomes will be assessed. The initial device will be removed and subsequent Airway management will then be undertaken with the Ambu AuraGain and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Assessment of device related trauma (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Insertion of Ambu AuraGain laryngeal mask airway and Igel (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Fibreoptic assessment with Ambu A-scope (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Ability to insert nasogastric tube (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Measurement of OLP (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Ability to perform positive pressure ventilation (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Assessment of ease of insertion (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Record number of manipulations (Device)

AuraGain and i-Gel

Active Comparator

In this arm of the crossover study, patients will undergo insertion of the Ambu AuraGain laryngeal mask airway and i-Gel and will undergo initial management with the Ambu AuraGain.. Primary and Secondary outcomes will be assessed. Airway management will then be undertaken with the iGel device and the same outcomes will be assessed.

Interventions with both devices will be

  1. Insertion of the laryngeal mask airway
  2. Assessment of ease of insertion
  3. Ability to perform positive pressure ventilation
  4. Measurement of OLP
  5. Fibreoptic assessment with Ambu A-scope
  6. Ability to insert nasogastric tube
  7. Record number of manipulations
  8. An assessment of device related trauma

干预措施: Assessment of device related trauma (Device)

结局指标

主要结局

oropharyngeal leak pressure

时间窗: Up to 3 minutes of successful insertion of each device

If successful ventilation is achieved with the device, the investigator will measure OLP by closing the Adjustable Pressure Limiting valve on the anaesthetic machine; with a fresh gas flow of 3l/min. Pressure at which equilibrium is reached will be deemed the OLP. Pressure will not be allowed to exceed 40cmH20.

次要结局

  • Ease of insertion(Up to 1 minute on each insertion attempt)
  • Number of attempts required to insert(Up to 1 minute of completion of total insertion attempts)
  • Evidence of Airway Trauma(Up to 1 minute after removal of first device used)
  • Adequacy of positive pressure ventilation(Up to 1 minute after successful insertion of device and after completion of outcome 4 measurement)
  • Fibreoptic assessment of device position(Up to 5 minutes after successful insertion and after completion of outcome 5 measurement)
  • Ability to insert nasogastric tube down gastric channel of device(Up to 5 minutes after successful insertion and after completion of outcome 6 measurement)
  • Intraoperative manipulations required(For up to 2 hours after induction of anaesthesia)