Validation of the WBPS - Wrist Blood Pressure System A Prospective, Method Comparison, Proof of Concept Study

CSEM Centre Suisse d'Electronique et de Microtechnique SA - Recherche et Developpement2 个研究点 分布在 1 个国家目标入组 50 人开始时间: 2016年10月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
50
试验地点
2
主要终点
Assessment of the accuracy of a novel algorithm to extract BP estimates from data acquired by the WinPack system, when compared to gold-standard BP devices, with respect to limits of agreement as stated by the IEEE 1708:2014 standard

研究概览

简要总结

The study will evaluate the accuracy of BP measurements by a new algorithm to be embedded to the existing CE-marked WinPack device, compared to two reference devices (one oscillometric and one volume-clamp device), in healthy subjects and patients undergoing a standardized physical-activity hypertensive protocol.

详细描述

Current standard for patient monitoring in acute care such as intensive care units (ICU), operating rooms (OR) or emergency rooms (ER) is the use of expensive, bulky and uncomfortable monitors.

Unfortunately, once discharged from the acute care setting and transferred to a general division for example (low acuity), patients are left with manual measurements of vital signs every 4 to 6 hours, depending on the available resources in nursing. Furthermore, these measurements reflect only the results of the time the measurement is taken (spot check). The combination of manual spot checks and lack of innovative medical technology in 60% of hospital care settings (all non-ICUs and non-OR) represents an important limitation to patient follow-up, with its associated risks and costs.

In particular, when referring to the monitoring of blood pressure (BP), current monitoring practice relies on the use of occlusive pneumatic cuffs inflated around the arm. Based on oscillometric technique, occlusive cuffs not only are uncomfortable, but they provide only intermittent measurements, with the inability to provide a full view of the BP regulation landscape. Therefore, un-detected hypertensive events can be omitted, putting the patient at risk.

The continuous, comfortable and portable measurement of BP in low acuity hospital departments is a major unsolved problem of modern healthcare systems with regards to patient monitoring and safety.

Winmedical (device manufacturer) commercializes a wearable multi-parametric modular medical system, WinPack, to deliver continuous real-time monitoring and recording of vital signs, such as electrocardiography (ECG), respiration rate, BP, blood oxygen saturation, body temperature, and posture, in low-acuity departments to increase the overall clinical performance. A new algorithm to process the raw physiological signals acquired by the WinPack to provide more accurate and more continuous BP measurement is now to be tested.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age >18 years
  • Informed Consent
  • Compliant and able to protocol manoeuvres

排除标准

  • Blood pressure difference between to arms >15mmHg
  • Current or history of tobacco smoking (for the last 5 years, > 5 cigarettes per day),
  • Current alcohol or drug abuse (>4 units per day),
  • Unstable cardiac condition (myocardial infarction < 1 week, pulmonary embolism, ventricular arrhythmia, decompensated heart failure)
  • Severe anaemia (haemoglobin <100g/l)
  • Oxygen therapy
  • Lesion or deficiency on left hand, preventing to perform handgrip exercises
  • Lesion or deficiency on both legs, preventing to perform leg extension exercise

研究组 & 干预措施

WinMedical WinPack device

Experimental

50 volunteers undergoing experimental protocol according to IEEE 1708:2014 standard

干预措施: WinMedical WinPack device (Device)

结局指标

主要结局

Assessment of the accuracy of a novel algorithm to extract BP estimates from data acquired by the WinPack system, when compared to gold-standard BP devices, with respect to limits of agreement as stated by the IEEE 1708:2014 standard

时间窗: 3 Months

次要结局

未报告次要终点

研究者

发起方
CSEM Centre Suisse d'Electronique et de Microtechnique SA - Recherche et Developpement
申办方类型
Industry
责任方
Sponsor

研究点 (2)

Loading locations...

相似试验