Usefulness of Ultra Low Dose Cardiac CT for Coronary Calcium Scoring and for Attenuation Correction of SPECT
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 105
- 试验地点
- 2
- 主要终点
- Agreement and Correlation of Coronary Artery Calcium Score Obtained From Ultra-low-dose and Standard CT
研究概览
简要总结
The aim of the study is to compare the CACS obtained from standard dose CT to the CACS obtained from ultra-low-dose scans. Additionally, the usefulness of ultra-low-dose CT for AC of myocardial perfusion SPECT will be assessed.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients referred for myocardial perfusion SPECT
- •Male and Female subjects ≥18 years of age,
- •Written informed consent
排除标准
- •Pregnancy or breast-feeding
- •CACS of 0 after inclusion of 10 patients with CACS 0
- •Stents or implanted cardiac devices (valves, pace makers, ICD)
- •Coronary artery bypass grafts
研究组 & 干预措施
Standard dose CT, Ultra-low dose CT
Standard dose non-contrast enhanced CT (clinically indicated) and Ultra low dose non-contrast enhanced CT (as part of the trial)
干预措施: Ultra low dose non-contrast enhanced CT (Device)
Standard dose CT, Ultra-low dose CT
Standard dose non-contrast enhanced CT (clinically indicated) and Ultra low dose non-contrast enhanced CT (as part of the trial)
干预措施: Standard dose non-contrast enhanced CT (Device)
结局指标
主要结局
Agreement and Correlation of Coronary Artery Calcium Score Obtained From Ultra-low-dose and Standard CT
时间窗: 1 days
CAC and BA limits of agreement between coronary artery calcium score obtained from ultra-low-dose and standard CT. Coronary artery calcium (CAC) is a measure for quantification of coronary artery calcification based on non-contrast enhanced CT, ranging from 0 (no calcifications) to infinite. It is an arbitrary unit. Increasing CAC means higher amounts of coronary artery calcifications and is associated with worse prognosis. Bland-Altman (BA) analysis is a statistical method to compare two modalities or techniques assessing the same measure. Limits of agreement is defined as +/- twice the standard deviation of the differences between the reference method and the new modality/technique. Broader limits of agreement mean less accurate results obtained by the new modality/techniqe, while a 0 BA limit of agreement would theoretically reflect perfect agreement.
Intra-class Correlation Coefficient Between Segmental Relative Tracer Uptake From SPECT Datasets Reconstructed With AC Maps Based on Ultra-Low-Dose and Standard Dose CT
时间窗: 1 day
For every patient, the CT images from 120 and 70 kVp-CT scans were used to create CTAC maps which were then used to reconstruct SPECT images, displayed as a 17-segment model polar plot with normalized percent tracer uptake given for every segment. Intra-class correlation was then applied to compare segmental relative tracer uptake. Analysis and the resulting correlation coefficient of 0.987 basically demonstrates interchangeability between the two datasets.
次要结局
未报告次要终点