Clinical study to research correlation of bone mineral density between Shimadzu DXA device and GE DXA device.

Oguchi Takeshi0 个研究点目标入组 130 人开始时间: 2019年5月31日最近更新:

试验速览

阶段
未知
状态
已完成
发起方
入组人数
130

研究概览

简要总结

We observed the correlation of BMD values obtained with the Shimadzu G4 and the GE Prodigy. And the correlation coefficients for the comparisons of these scanners for the participants' BMD values were 0.981-0.959. Mean errors were 0.81%-0.03%. As a result, the correlation coefficients were high. CV values of BMD values measured on the Shimadzu G4 were 1.01% (lumbar spine) and 1.28-1.44% (femoral neck ). We also observed high reproducibility for BMD values measured on the Shimadzu G4.

研究设计

研究类型
Interventional

入排标准

年龄范围
>= 20age old 至 ot applicable(—)
性别
All

入选标准

  • 1)The person whose is age 20 years old or more.(Regardless of gender)
  • 2)The person who can understand this study and sign the consent form of this study by authentic signature.

排除标准

  • 1)The persons who is not suitable for the measurement of DXA, for example, the person with the vertebral fracture past or proximal femur fracture past in measurement range(L1-L4), the person who has salient scoliosis, and who has fixtures in the body such as metal pin, wire, screw, plate, implant.
  • 2)Pregnant or the persons who may be pregnant.
  • 3)The persons who the investigator or subinvestigator determine unsuitable for participation.