Fast VT Episodes Are Terminated by ATP One Shot

Biotronik SE & Co. KG27 个研究点 分布在 4 个国家目标入组 194 人开始时间: 2007年11月最近更新:
适应症

试验速览

阶段
4 期
状态
已完成
入组人数
194
试验地点
27
主要终点
Time to the first adequate device shock

研究概览

简要总结

Many arrhythmias detected in the ventricular fibrillation (VF)-zone picture monomorphic ventricular tachycardia (VT) and, hence, could be terminated by antitachycardia pacing (ATP) treatment. Advantages of successful ATP are the painlessness termination and the shortened duration of arrhythmia. The ATP One Shot algorithm is integrated in the latest family of implantable cardioverter defibrillators (ICDs) from BIOTRONIK (Lumax). It allows a single delivery of ATP before charging capacitors to terminate lethal arrhythmia by painful shock.

The present study evaluates the efficacy of the ATP One Shot algorithm for the termination of fast VT episodes. 200 patients with secondary prophylactic ICD indication will be followed for 18 months. Spontaneous episodes detected in the VF-zone of the ICD will be evaluated with regard to cycle length, episode duration and course of device therapy.

详细描述

ICDs provide different detection zones for the therapy of VT and VF. An established form of therapy in the VT-zone is antitachycardia pacing (ATP). It is an effective, safe and painless method to terminate episodes of slow VT with a cycle length > 300 ms. Episodes detected in the VF zone will result in the delivery of a high-energy shock to terminate life-threatening VF. Many arrhythmia detected in the VF-zone picture monomorphic VT and could be easily terminated by ATP. Advantages of a successful termination by ATP therapy would be the painlessness and the shortened duration of the episode, as there would be no need to load shock capacitors. Fast VT are often hemodynamically poorly tolerated by the patient and should be terminated within very short time. The ATP One Shot algorithm is integrated in the latest family of ICDs from BIOTRONIK (Lumax) to allow a single delivery of ATP before charging capacitors.

The main objective of the study is the assessment of the efficacy of the ATP One Shot algorithm for the termination of fast VT episodes. To this end, spontaneous episodes detected in the VF-zone of the ICD are evaluated with regard to cycle length, episode duration and course of device therapy. In the context of the study, a confirmatory (hypothesis-testing) primary problem is investigated. The goal is to measure the time to first adequate shock therapy for episodes detected in the VF-zone of the device and to analyze the hazard ratio of the standard ICD setting compared to a therapy with the ATP One Shot algorithm.

The clinical project is conducted as randomized prospective multicenter study. The chronological order and the scope of the follow-ups meet the medical standard according to the international guidelines. A center-based, stratified block randomization will be performed. In approximately 50% of the patients the ATP One Shot algorithm will be activated, while the other 50% will be randomized into the control group.

A total of about 20 European investigational sites will participate in the study. Each center should enroll about 10 patients who will be followed for 18 months.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • ICD indication for reason of survived cardiac arrest, ventricular tachycardia or syncope

排除标准

  • ICD indication for reason of non-sustained VT post MI, positive family history for SCD (Brugada-, long/short QT-Syndrome), hypertrophic cardiomyopathy, primary prophylactic indication (e.g. SCD-HeFT, Madit II)
  • Patients not capable of participating in the follow-ups
  • Minors and pregnant women
  • Patients who are already enrolled in another study

结局指标

主要结局

Time to the first adequate device shock

时间窗: 18 months

次要结局

  • Quality of Life(18 months)
  • Duration of Episodes(18 months)