Evaluation of Blood Glucose Meter Systems - Contour® PLUS Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 106
- 试验地点
- 2
- 主要终点
- MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
研究概览
简要总结
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
详细描述
The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems were evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Males and females, 18 years of age or older
- •Willing to complete all study procedures
排除标准
- •Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
- •Hemophilia or any other bleeding disorder
- •Pregnancy
结局指标
主要结局
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range
时间窗: 8 hours
Using the overall Blood Glucose (BG) range (27 to 460 mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.
次要结局
- MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range(<=80 mg/dL)(8 hours)
- MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)(8 hours)