跳至主要内容
临床试验/NCT02697214
NCT02697214已完成不适用

Accuracy of Commercially Available Heart Rate Monitors: A Prospective Study

The Cleveland Clinic0 个研究点目标入组 50 人开始时间: 2015年8月最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
50
主要终点
Measuring the Accuracy of Commercial Heart Rate Monitors.

研究概览

简要总结

Over the last two decades, there has been a proliferation of commercially available heart rate monitors. Recognizing that elite athletes often use heart rate to monitor training and assess aerobic fitness, fitness companies have offered a variety of heart rate monitoring systems to the general public. Recently, there has been a move from monitors that rely on chest straps to measure electrical activity toward more convenient, wrist-worn monitors that employ optical sensing technology similar to that used for pulse oximetry. While the accuracy of chest strap monitors has been assessed in a variety of studies, there is no data concerning the accuracy of wrist-worn heart rate monitors. Assessment of the monitors' accuracy is important both for the subjects who rely upon these monitors to guide their athletic activity and for the physicians to whom these individuals report their heart rate readings.

详细描述

Physicians have long-used heart rate as an index of aerobic fitness and cardiovascular health. Resting heart rate has been linked to longevity and freedom from cardiovascular morbidity, and heart rate during exercise serves as a measure of aerobic capacity. In addition, heart rate recovery after exercise carries implications for an individuals' health.

Like physicians, coaches and athletic trainers employ heart rate measurement on a daily basis. In the athletic setting, heart rate during exercise is used as an indication of aerobic exertion. Elite athletes and their coaches often design workouts based upon achieved and targeted heart rates.

Once the province of physicians and elite athletes, heart rate monitoring has become widespread among the general public. In the nineteen eighties, fitness companies added heart rate monitors to their product lines. Systems employing a chest strap to monitor electrical activity, telemetry and a wrist-borne receiver became popular, and several controlled, scientific studies confirmed their accuracy.

More recently, manufacturers have marketed a new class of heart rate monitors that consist solely of wristwatch-style devices. These heart rate monitors use optical sensing technology to measure heart rate. While they offer convenience, their accuracy, particularly during exercise, is uncertain. A recent article in USA Today suggests that these wrist-worn monitors fail to provide accurate readings during exercise; however, to date there has been no rigorous scientific inquiry addressing this question.

The purpose of this study is to assess the accuracy of four popular, commercially available wrist-worn heart rate monitors under conditions of varying physical exertion.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Factorial
主要目的
Screening
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age > 18 years
  • Able and willing to exercise (walk/jog) for a total of fifteen minutes

排除标准

  • Health issues that preclude or contraindicate walking and/or jogging, including cardiovascular, orthopedic, pulmonary and other conditions
  • Presence of a cardiac pacemaker
  • Known cardiovascular disease
  • Known heart rhythm disorders
  • Use of Beta-blockers or antiarrhythmic medications
  • Tattoos around the wrist area

研究组 & 干预措施

Apple Watch

Active Comparator

Apple Watch heart rate monitoring device worn with standard ECG, Polar H7 chest monitor and one other wrist monitoring device .

干预措施: Apple Watch (Device)

Fitbit Charge HR

Active Comparator

Fitbit Charge HR heart rate monitoring device worn with standard ECG, Polar H7 chest monitor and one other wrist monitoring device.

干预措施: Fitbit Charge HR (Device)

Mio Fuse

Active Comparator

Mio Fuse heart rate monitoring device worn with standard ECG, Polar H7 chest monitor and one other wrist monitoring device.

干预措施: Mio Fuse (Device)

Basis Peak

Active Comparator

Basis Peak heart rate monitoring device worn with standard ECG, Polar H7 chest monitor and one other wrist monitoring device.

干预措施: Basis Peak (Device)

结局指标

主要结局

Measuring the Accuracy of Commercial Heart Rate Monitors.

时间窗: 20 minutes

The purpose of this study is to assess the accuracy of four popular, commercially available wrist-worn heart rate monitors compared to the current gold standard of a ECG using Lin's Concordance Correlation Coefficient.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Marc Gillinov, MD

MD

The Cleveland Clinic

相似试验