Clinical Validation of the Aptitude Medical Systems Metrix COVID-19 Test for Detection of SARS-CoV-2 in Point-of-Care and At-Home (OTC)/Non-Laboratory Settings

Aptitude Medical Systems6 个研究点 分布在 1 个国家目标入组 1,000 人开始时间: 2024年10月17日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
发起方
入组人数
1,000
试验地点
6
主要终点
Qualitative detection of RNA from SARS-CoV-2 and Measurement of Positive/Negative Percent Agreement with FDA-cleared comparator test

研究概览

简要总结

The Metrix COVID Test will be evaluated for use in Point-of-Care and Non-Laboratory settings in a home testing environment utilizing the clinical study design described herein. The study will take place in simulated home environments which will be set up within or near active clinical settings (e.g., urgent care facilities). This will be a prospective study conducted at three or more investigational sites located within the United States for the clinical validation of the Metrix COVID Test for the detection of SARS-CoV-2 AN swab samples. Additional sites may be added to the study in order to meet minimum subject/sample enrollment requirements and geographic prevalence of respiratory virus infections. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The primary comparator for the study will be an FDA-cleared assay for the detection of SARS-CoV-2.

详细描述

The Metrix COVID Test will be evaluated for OTC use in a home use testing environment utilizing the clinical study design described herein. The study will take place in simulated home environments which will be set up within or near active clinical settings (e.g., urgent care facilities). This will be a prospective study conducted at three or more investigational sites located within the United States for the clinical validation of the Metrix COVID Test for the differential detection of SARS-CoV-2 in AN swab samples. Additional sites may be added to the study in order to meet minimum subject/sample enrollment requirements and geographic prevalence of respiratory virus infections. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The primary comparator for the study will be an FDA-cleared assay for detection of SARS-CoV-2.

The candidate device, Metrix COVID Test, will be evaluated with AN swab sample types. Each Metrix COVID Test kit contains all supplies needed to test AN swab samples, including a corresponding QRI for detailed instructions.

This OTC study will take place in simulated home environments which will be set up within or near clinical settings (e.g., urgent care facilities). The study will enroll symptomatic subjects only. Candidate samples will be self-collected by participants (or collected by a guardian for participants under the age of 14 years) and comparator samples will be collected by a healthcare practitioner with informed consent and Institutional Review Board (IRB) approval. Each sample will be coded for confidentiality.

For each participant, a comparator NP swab sample will be obtained by a healthcare practitioner after testing with the candidate device. This NP swab will be tested with the FDA-cleared SARS-CoV-2 assay according to the IFU and will serve as the primary comparator for swab samples. An additional AN swab sample will be collected and stored frozen and dry for later analysis to support future research validation studies.

No results from the investigational test or comparator testing will be used for clinical diagnosis, management of study participants, or to make treatment decisions.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
2 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Participant or guardian understands and is able and willing to provide written informed consent, and assent where applicable, prior to study enrollment.
  • Male or female aged 2 years or older
  • Participant is currently exhibiting fever, or one or more symptoms of respiratory tract infection (such as, but not limited to, chills, cough, shortness of breath or difficulty breathing, fatigue, new loss of taste or smell, sore throat, congestion or runny nose, nausea or vomiting or diarrhea). Participant must still be exhibiting symptoms on the day of sample collection.
  • Participant or guardian agrees to read, and is able to read with understanding, the AN swab QRI prior to beginning the operation of each of the Metrix COVID-19 Test.
  • Participant or guardian is able and willing to contribute the required swab samples for testing and understands and is able and willing to sign the study informed consent.

排除标准

  • Participant does not understand and/or is not able and willing to sign the study informed consent and/or assent.
  • Participant or guardian is not able to comply with nasal swab collection requirements following the QRI.
  • Participant has previously provided a sample for the study.
  • Participant is not able to tolerate sample collection.
  • Participant is currently undergoing antiviral treatment such as baloxavir marboxil (trade name Xofluza®), oseltamivir (Tamiflu®), zanamivir (Relenza®), and peramivir (Rapivab®).
  • Participants currently undergoing treatment and/or within the past thirty (30) days with prescription medication to treat novel Coronavirus SARS-CoV-2 infection, which may include but is not limited to Remdesivir (Veklury), Paxlovid, molnupiravir or receiving convalescent plasma therapy for SARS-CoV-
  • Participants who have had a nasal wash or aspirate as part of their standard of care treatment on day of study visit prior to the study sample collection.
  • Participants who have had recent craniofacial injury or surgery, including to correct deviation of the nasal septum, within the previous six (6) months.
  • Participants who do not understand/read the English language.

结局指标

主要结局

Qualitative detection of RNA from SARS-CoV-2 and Measurement of Positive/Negative Percent Agreement with FDA-cleared comparator test

时间窗: 25 minutes

Qualitative detection of RNA from SARS-CoV-2 in unprocessed AN swab samples. Primary objective of this study is to evaluate the performance (positive percent agreement, negative percent agreement and 95% confidence intervals for upper and lower bounds) of the Metrix COVID/Flu Test for the qualitative detection of SARS-CoV-2 viral RNA using AN swab samples. Comparator testing will be performed to determine the status of each sample for comparison to results produced by the candidate test. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The primary comparator for the study will be an FDA-cleared assay that detects SARS-CoV-2.