CareTaker Self-Calibration Validation

CareTaker Medical LLC0 个研究点目标入组 140 人开始时间: 2016年3月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
140
主要终点
Performance comparison

研究概览

简要总结

The goal of the study is to test the capability of the CareTaker monitor to calibrate its continuous blood pressure readings independently. Currently the device requires another approved blood pressure monitor to provide a starting calibration.

The new control module enables the CareTaker to perform a pressure sweep of the internal pressure in the finger cuff. The resulting data is analyzed using a combination of pulse analysis/oscillometry approach.

详细描述

The specific objective of this project was to demonstrate the CareTaker's ability to perform self-calibration to the same performance standard as its predicate device. That device demonstrated efficacy by showing substantial equivalence to the performance of classical sphygmomanometry using a Riva-Rocci/Korotkoff (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg across 3 paired readings from at least 85 patients, for a total of at least 255 paired readings.

The study was performed at the University of Virginia Medical Center in Charlottesville, Virginia. UVA's Institutional Review Board approved (#18686) and supervised the study for subjects who were University of Virginia hospital patients and staff > 18 years of age and who were able to give verbal consent.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • > 18 years of age, who are able and willing to participate and have given verbal assent

排除标准

  • Unable to give verbal assent
  • <18 years of age
  • No or poor finger pulse, as determined through visual inspection for ischemic hands

结局指标

主要结局

Performance comparison

时间窗: 8 minutes for 3 paired readings

Comparison with performance of classical sphygmomanometry using a Riva-Rocci/Korotkov (RRK) measurement on the brachial artery, meeting a bias of at least 5 mmHg and a standard deviation of 8 mmHg