The Effect Of Auricle Position On Body Temperature Measurements Made With Tympanic Membrane Thermometer In Adults
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 143
- 主要终点
- outcome1
研究概览
简要总结
Objective: The aim of this study was to investigate the effect of auricle position on body temperature measurements made with tympanic membrane thermometer in adult patients.
Methods: The study was conducted with a quasi-experimental design with a pre-test and a post-test. A total of 143 patients who fit the inclusion criteria of the study were included in the sample. The research carried out between 1 November - 31 December 2016 in the emergency department. Body temperature measurements were carried out on the patients in the study firstly by not changing the position of the auricle and then repeated after a minute, this time by changing the position of the auricle.
Results: The difference between the measurement values found in two different positions was 0.31 0C, and the Bland Altman plot showed that the differences were distributed systematically around the value 0.31. The difference between the lengths of measurements performed on two separate positions was detected to be 1.08 second. During body temperature measurements performed through giving position to auricle, 59.2 % of the patients never felt discomfort.
Conclusions: It was found in the comparison of two positions that there was a significant difference between the tympanic membrane thermometer measurements made by positioning the auricle and those without positioning.
详细描述
Introduction One of the most common practices in clinical investigations is the measurement of a patient's body temperature (1-2,3). Body temperature measurement has a significant place in evaluation of clinical pictures and clinical monitoring (4). The issue at hand is that various temperature measurement modalities yield different results (2-4).
Although the accuracy of temperatures measured by infrared tympanic thermometers is a controversial issue among people interested in the subject, these thermometers have many advantages; they measure the temperature fast (1-2 s), and they are easy to use and non-invasiveness (3). Compared to the armpit or oral temperatures, the tympanic temperature reflects the core temperature more accurately which was demonstrated by comparing the temperature measurements made in the pulmonary artery and esophagus (5-6).
Methods Aim The aim of this study investigate the effects of auricle position on body temperature values in tympanic membrane thermometer measurements performed on adult patients. The hypothesis was "Position of the auricle during measurement of tympanic membrane temperature will affect the reading".
Design The study was conducted with a quasi-experimental design with a pre-test and a post-test.
The Study Participants This cross-sectional study surveyed 143 patients from one of emergency department in Turkey. The patients who were 18 years old or older, conscious, without communication issues, volunteers, not diagnosed with otitis media, who did not have a fever. Because otitis media and fever cause to potential sources of bias. A sample of the study with 143 patients who meeting these criteria and agreed to participate in the study was generated.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Other
- 盲法
- None
盲法说明
A total of 143 patients who fit the inclusion criteria of the study were included in the sample. The research carried out between 1 November - 31 December 2016 in the emergency department. Body temperature measurements were carried out on the patients in the study firstly by not changing the position of the auricle and then repeated after a minute, this time by changing the position of the auricle.
入排标准
- 年龄范围
- 18 Years 至 70 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •The patients who were 18 years old or older, conscious, without communication issues, volunteers, not diagnosed with otitis media, who did not have a fever.
排除标准
- •havin fever, Being under the age of 18, not volunteers.
结局指标
主要结局
outcome1
时间窗: NOVEMBER 2016 JANUARY 2017
It was found in the comparison of two positions that there was a significant difference between the tympanic membrane thermometer measurements made by positioning the auricle and those without positioning