临床试验/NCT02125331
NCT02125331终止不适用

Measurement of NonInvasive Blood Pressure With DINAMAP SuperSTAT and Datex-Ohmeda With Intra-arterial Blood Pressure in Neonates Through Adults and Special Populations

GE Healthcare4 个研究点 分布在 2 个国家目标入组 83 人开始时间: 2014年4月最近更新:
适应症

试验速览

阶段
不适用
状态
终止
发起方
GE Healthcare
入组人数
83
试验地点
4
主要终点
Number of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements

研究概览

简要总结

This study is required to demonstrate that the non-invasive blood pressure (NIBP) measurement algorithms on two commercially available multifunction hemodynamic acquisition modules, the Patient Data Module (PDM) equipped with the DINAMAP® SUPERSTAT algorithm ("PDM-SUPERSTAT") and the Patient Side Module (PSM) equipped with Datex-Ohmeda GE algorithm ("PSM-Datex-Ohmeda"), provide accurate NIBP measurements in accordance with the guidelines provided by the most recent International Organization for Standardization (ISO) 81060-2:2013.

Notably, the new ISO 81060-2:2013 supersedes the previous 2009 version of this standard, and this study is being conducted in accordance with the most recent applicable standards.

This study will assess neonate, infant, children, pediatric, adolescent, and adult patients as well as adults with chronic atrial fibrillation in a population that requires non-emergent heart catheterization.

This is a study conducted to satisfy both standards for two NIBP devices and, as such, does not require a statistical hypothesis.

详细描述

This study is required to demonstrate that the non-invasive blood pressure (NIBP) measurement algorithms on two commercially available multifunction hemodynamic acquisition modules, the Patient Data Module (PDM) equipped with the DINAMAP® SUPERSTAT algorithm ("PDM-SUPERSTAT") and the Patient Side Module (PSM) equipped with Datex-Ohmeda GE algorithm ("PSM-Datex-Ohmeda"), provide accurate NIBP measurements in accordance with the guidelines provided by the most recent International Organization for Standardization (ISO) 81060-2:2013.

Notably, the new ISO 81060-2:2013 supersedes the previous 2009 version of this standard, and this study is being conducted in accordance with the most recent applicable standards.

This study will assess neonate, infant, children, pediatric, adolescent, and adult patients as well as adults with chronic atrial fibrillation in a population that requires non-emergent heart catheterization.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • 未提供

排除标准

  • 未提供

结局指标

主要结局

Number of Participants With Accurate NIBP Measurements as Confirmed by IBP Measurements

时间窗: 60 minutes

Comparison of the following measurements: non-invasive blood pressure to invasive blood pressure. This is done with a beat to beat comparison. Each pulse from the analyzed time interval (non-invasive blood pressure determination) provides a systolic, MAP and diastolic value and that is compared to the waveform collected at the same timepoint for the invasive BP (measurement). Criteria is defined in ANSI/AAMI/ISO 81060-2 NIBP Standard. Mean \& SD of the IBP derived from the IBP wave recordings during a NIBP determination will be calculated; and the range of reference IBP will be determined as mean ±1 standard deviation (SD) of the IBPs. The same method will be used for both systolic and diastolic blood pressures. Data may be excluded for analysis when the range of invasive systolic BP is wider than 20 mmHg or when the range of invasive diastolic BP is wider than 12 mmHg, as such that analysis is performed in accordance with ISO 81060-2:2013.

次要结局

未报告次要终点

研究者

发起方
GE Healthcare
申办方类型
Industry
责任方
Sponsor