Evaluating the Accuracy of a Neonatal Temperature Monitor in Comparison With Existing Temperature Monitors
试验速览
- 阶段
- 不适用
- 入组人数
- 150
- 试验地点
- 2
- 主要终点
- Temperature measurement accuracy
研究概览
简要总结
A team of researchers at Rice University and Queen Elizabeth Central Hospital (QECH) are working to develop a low-cost temperature sensor that can continuously monitor an infant's temperature (NTM). This robust, low-cost device will allow for the individualized monitoring of each infant with alerts for hypo and hyperthermia. A reusable band placed around the infant's abdomen to hold the temperature sensor will eliminate disposable components. This study will assess the accuracy of this novel device against a gold standard (Philips Intellivue patient monitor) and up to two existing devices (Bempu and Thermospot).
详细描述
This study will assess the accuracy of this novel device against a commercially available patient monitor (Philips Intellivue MP30). Results will be compared to that of two existing devices designed for low resource settings: Bempu, a temperature monitor placed around an infant's wrist; and Thermospot, a sticker that changes color with temperature.
This study will enroll up to 150 infants at Queen Elizabeth Central Hospital. Up to 75 subjects will be collected from the neonatal ward and 75 subjects from the Kangaroo Mother Care (KMC) ward. This sample size will ensure that the investigators are able to collect sufficient data from infants with a range of gestational ages, weights, temperatures, and treatment locations (ie open cot, radiant warmer). The data from this study can be used to calculate the sample size needed in a larger study to evaluate changes in outcomes and nurse behavior related to the different temperature monitors (NTM, Bempu, Thermospot).
During the trial, the following steps will be taken:
A trained study nurse will assess the subject for clinical complications before attaching the temperature monitoring devices.
- A trained study nurse or clinician will attach the test temperature monitoring devices to the infant. A trained research assistant from Rice University or from the Biomedical Engineering Department at Malawi Polytechnic will observe all procedures and will notify the nurse of any observed errors so they may be corrected. They will also be able to answer any technical questions from the nurse. If there is any concern for the comfort or skin of the infant, only one device may be placed on at a time (NTM, Bempu, ThermoSpot).
- The trained study nurse will attach the temperature probe from the commercial gold standard as well as provide any other care needed.
- A research assistant will use a laptop to collect data from the continuous temperature monitors (NTM and gold standard). The nurse will record the temperature readings every hour from all monitors on paper forms along with the standard of care.
- Each time an alarm sounds from any device (Bempu and the commercial gold standard have audible alarms), the nurse will record all temperature values including a reading taken with the standard of care method. If the subject is found to be hyper/hypothermic, the nurse will respond with the appropriate standard of care.
- Temperature monitoring will continue for up to 3 days. The research assistant may ask the nurse to remove and reapply the temperature probes during this period.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Device Feasibility
- 盲法
- None
入排标准
- 年龄范围
- — 至 28 Days(Child)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •The subject is currently being treated at QECH in the neonatal or KMC ward.
- •Study devices are available for use.
- •The subject's caregiver has provided informed consent for their child to participate.
排除标准
- •At the clinician's discretion for any reason including, but not limited to:
- •potential for skin irritation
- •other condition that may preclude use of the temperature belt
- •concurrent treatments that may require increased patient care
结局指标
主要结局
Temperature measurement accuracy
时间窗: <72 hours
Continuously measured temperature for test devices compared against Philips Intellivue patient monitor
次要结局
未报告次要终点