Clinical Evaluation of the iMOST Analyzer - Matrix Study Comparing Capillary and Venous Whole Blood
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 60
- 试验地点
- 1
- 主要终点
- Method Comparison of CBC Results
研究概览
简要总结
This is a method comparison study for an in vitro diagnostic device. The device is a point of care automated hematology analyzer which measures CBC parameters with a small drop of venous or capillary, e.g., fingerstick blood. The study compared the CBC results from the test device, the Essenlix iMOST X-1 to the predicate laboratory based Horiba Pentra 60C+ analyzer. The CBC parameters were hemoglobin, White Blood Cells (WBC) and the WBC Differential for Granulocytes, Neutrophils, Lymphocytes and Monocytes. The study subjects were patient donors (age 18 and older) attending the hospital clinic from whom Informed Consent had been received.
详细描述
These studies demonstrate that the iMOST X-1 system yields equivalent results to the Horiba predicate system when using either capillary and venous blood specimens. Furthermore, there are three alternative methods of capillary blood sampling for the iMOST and all have direct applicability to the iMOST system. As has been published in the peer reviewed literature, different fingerstick sample collection methods yield slightly different CBC results, specifically the WBC and the WBC Differential. The iMOST employs an integrated software algorithm with which the use can employ site specific calibration constants to correctly match the selected capillary blood specimen to the venous specimen for the WBC test results. All three methods of blood collection are thus described in the "Instructions for Use".
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •all clinic attendees
排除标准
- •unable to provide a venous and fingerstick blood specimen
结局指标
主要结局
Method Comparison of CBC Results
时间窗: October- December 2020
Blood Diagnostic Tests for CBC
次要结局
未报告次要终点