临床试验/NCT06007196
NCT06007196尚未招募不适用

Performance Evaluation of Non-invasive Continuous Hemodynamic Measurement From Task Force(R) CORE Versus Invasive Reference

Medical University of Vienna1 个研究点 分布在 1 个国家目标入组 70 人开始时间: 2023年9月1日最近更新:
适应症

试验速览

阶段
不适用
状态
尚未招募
入组人数
70
试验地点
1
主要终点
Difference in intraoperative cardiac output measurements between Task Force CORE compared with thermodilution during surgery

研究概览

简要总结

Validation of hemodynamic measurements using the non-invasive Task Force Core(R) and Task Force Cardio(R) device with invasive hemodynamic measurements using the thermodilution technique (PICCO device).

详细描述

The investigational device, Task Force® CORE (TFC, CNSystems Medizintechnik GmbH, Graz, Austria) together with the Task Force® CARDIO (TFCARDIO, CNSystems Medizintechnik GmbH, Graz, Austria) medical device software, is a new CE-marked medical device for continuous non-invasive determination of blood pressure (BP), cardiac output (CO) and derived parameters based on the well established CNAP® (continuous non-invasive arterial pressure) technology by CNSystems. The primary aim of this prospective, method comparison, open, multi-center study is data acquisition for the performance evaluation of the TFC in comparison with invasive reference methods (i.e. transpulmonary thermodilution, invasive blood pressure with radial artery catheter). The setting of the study will be the Operating Room (OR). This setting allows the comparison of the investigational device's cardiac output (CO) to invasive CO as well as of BP and derived hemodynamic parameters like Stroke Volume (SV), and Systemic Vascular Resistance (SVR). Measurements will be performed during the perioperative period, starting 15 minutes after the first skin incision.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adult patients (patients aged at least 18 years)
  • Patients undergoing major abdominal surgery who require invasive extended hemodynamic monitoring (trans pulmonary thermodilution via PiCCO and invasive blood pressure monitoring with radial artery catheter) according to clinical standard of care
  • Patients with intact perfusion of both hands evidenced by a positive Allen's test
  • Patients given written informed consent to participate in the study

排除标准

  • Patients with vascular impairments at the sites of non-invasive blood pressure measurement (fingers or upper arms)
  • Patients with significant edema in the fingers
  • Patients with ventricular assist devices or extracorporeal life-supporting devices
  • Subjects not passing the Allen's test for both hands
  • Patients who obtain vasoactive medications on the same arm as used for upper-arm Task Force blood pressure measurements
  • Patients with a large lateral difference in blood pressure (> 15mmHg for systolic blood pressure and/or >10 mmHg for diastolic blood pressure) or with same arm measurement differences > 10 mmHg in systolic or diastolic blood pressure during assessment of lateral differences
  • Patients who, in addition to invasive blood pressure monitoring, require non-invasive blood pressure measurements unrelated to the investigational device

结局指标

主要结局

Difference in intraoperative cardiac output measurements between Task Force CORE compared with thermodilution during surgery

时间窗: For the duration of surgery

Intraoperative performance of cardiac output values derived by the Task Force CORE system compared with cardiac output measurements derived by thermodilution measurements.

次要结局

  • Differences in intraoperative hemodynamic measurements (stroke volume in ml) between Task Force CORE compared with thermodilution during surgery(For the duration of surgery)
  • Intraoperative continuous Cardioc Output measurements(For the duration of surgery)
  • Differences in intraoperative hemodynamic measurements (stroke volume variation in %) between Task Force CORE compared with thermodilution during surgery(For the duration of surgery)
  • Differences in intraoperative hemodynamic measurements (blood pressure in mmHg) between Task Force CORE compared with thermodilution during surgery(For the duration of surgery)