跳至主要内容
临床试验/NCT02773901
NCT02773901Unknown不适用

Comparison of the Non-invasive ICP HeadSense Monitor vs Lumbar CSF Pressure Measurement

HeadSense Medical3 个研究点 分布在 3 个国家目标入组 60 人开始时间: 2015年11月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
发起方
入组人数
60
试验地点
3
主要终点
Number of ICP values obtained by the HeadSense device that correlate to ICP using current standards

研究概览

简要总结

In current practice patients with a suspected increase of intracranial pressure (ICP) will undergo a lumbar puncture with measurement of cerebral spinal fluid (CSF) pressure (as a marker for ICP). A lumbar puncture is an invasive and sometimes painful procedure. Using a new type of ICP monitor (HeadSense) it is possible to measure ICP non-invasively through an acoustic signal

详细描述

In this study we will prospectively collect relevant clinical data on 60 neurological patients, who are being referred for an elective lumbar puncture (eg. suspected multiple sclerosis, raised intracranial hypertension, etc). Each enrolled patient will be monitored in parallel to the lumbar puncture with the HeadSense ICP monitor. Subjects who meet the study inclusion and exclusion criteria will be enrolled in the study.

Step 1: 10 minute measurement of HeadSense monitor. Once the patient is inclined with his upper body 30 degrees to the bed the pressure values, the clinical procedure can begin. The device must be preset in a continuous monitoring mode. The continuous monitoring allows a loop of measurements for an unlimited time in a rate of four measurements per minute.

Step 2: 10 minute measurement of HeadSense monitor in supine position.

Step 3: Patient will be placed in a left lateral position. The lumbar puncture will be done and CSF pressure will be measured following standard procedures.

Step 4: 10 minute measurement of HeadSense monitor in supine position.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Male or Female in the age range of 18 years and older
  • All patients who are referred for elective or subacute lumbar puncture including CSF pressure measurement

排除标准

  • Local infection in the ear
  • Pregnant or lactating women
  • Cervical spine stenose
  • Arnold-Chiari malformations
  • Aqueductal stenosis or other changes causing an uneven CSF pressure between different compartments
  • Mass lesions
  • Current or previous craniotomy or craniectomies
  • Suspicion of meningitis or encephalitis

研究组 & 干预措施

HS-1000 recording

Experimental

ICP monitoring will be done with the HS-1000 to compare to CSF measurement from lumbar puncture (per clinical protocol without any change in the patient's management). HS-1000 ICP monitoring intervals last 10 minutes.

干预措施: HS-1000 (Device)

结局指标

主要结局

Number of ICP values obtained by the HeadSense device that correlate to ICP using current standards

时间窗: 10 minutes

次要结局

  • Incidence of Adverse events (AEs)(48 hours from the end of monitoring with the HS-1000)

研究者

发起方
HeadSense Medical
申办方类型
Industry
责任方
Sponsor

研究点 (3)

Loading locations...

相似试验

Comparison of the Non-invasive ICP HeadSense Monitor... | 临床试验