B40 Non-Invasive Blood Pressure Equivalency Study

GE Healthcare2 个研究点 分布在 1 个国家目标入组 66 人开始时间: 2012年3月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
GE Healthcare
入组人数
66
试验地点
2
主要终点
Difference Between the B40 Monitor and Reference Device DINAMAP PRO1000 on NIBP Measurements

研究概览

简要总结

The purpose of this study is to demonstrate equivalency of the GE Healthcare B40 Patient Monitor noninvasive blood pressure (NIBP) parameter compared to the reference device.

详细描述

Automated non-invasive blood pressure devices have a long history. These devices first made their appearance in the mid-1970's. Since that time, there have been many improvements in the performance of these devices. Specifically, when making changes to the algorithms that control these devices, the goals are: to improve accuracy and precision in the blood pressure (BP) estimates, to reduce the time needed to determine the BP estimates, to increase comfort for the patient, and to work through artifacts and arrhythmias that are often present in the clinical environment.

The purpose of this study is to demonstrate equivalency of the GE Healthcare B40 Patient Monitor noninvasive blood pressure (NIBP) parameter compared to the reference device.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
3 Years 至 75 Years(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Ability to provide written informed consent or have a legally authorized representative including a parent or guardian provide written informed consent
  • Within the age range from older than 3 years old to 75 years old (adult and child)
  • Ability to maintain proper placement of cuff and/or probe and/or sensor

排除标准

  • Neonates (ages 0-29 days) and infants (30 days to 3 years old) shall be excluded
  • Any subject deemed too unstable, by clinician's discretion, or for another reason deemed unacceptable for study by the clinician
  • Known dysrhythmias (when applicable)
  • Known disease state that compromises circulation to the extremity
  • Vigorous exercise prior to participating in the study
  • Excessive movement or excitability causing false values or no determinations
  • Known allergy to latex when latex products will be in contact with subject
  • Pregnant females

结局指标

主要结局

Difference Between the B40 Monitor and Reference Device DINAMAP PRO1000 on NIBP Measurements

时间窗: End of each blood pressure reading

The primary endpoints are the difference in systolic, diastolic and mean blood pressure values between the B40 Monitor and PRO1000. The mean of the difference should be no more than 5mmHg and the standard deviation of the difference should be no more than 8mmHg per the AAMI SP-10 standard.