Multisite Precision Study for the SpinChip High-sensitivity Cardiac Troponin I (SpinChip Hs-cTnI) Test

SpinChip Diagnostics ASA3 个研究点 分布在 1 个国家目标入组 37 人开始时间: 2025年2月3日最近更新:

试验速览

阶段
不适用
状态
已完成
入组人数
37
试验地点
3
主要终点
Precision of the SpinChip hs-cTnI test using Li-heparin plasma and whole blood samples

研究概览

简要总结

During a heart attack, the protein troponin I is released from the heart muscle into the bloodstream. Measurements of cardiac troponin in blood are used as an aid in the diagnosis of heart attack. The SpinChip hs-cTnI test is a new high-sensitive test for measuring the amount of cardiac troponin I in the bloodstream as an aid in the diagnosis of heart attack.

The purpose of this study is to establish the precision of the SpinChip hs-cTnI test when using Li-Heparin plasma and whole blood samples in the hands of the intended user in the intended use setting at multiple sites.

详细描述

Cardiac troponins are widely used as a biomarker to aid in the diagnosis of acute myocardial infarction (AMI). These structural proteins are essential in regulating contraction in cardiac muscle cells, and they are sensitive and specific biochemical markers of myocardial damage.

During a heart attack, cardiac muscle cells are injured and release the cardiac marker troponin I (cTnI) into the bloodstream. High-sensitive troponin tests may detect the increase of cardiac troponin in blood within hours after the symptoms of a heart attack has started.

The SpinChip high-sensitivity cardiac troponin I (hs-cTnI) test is a new high-sensitive test for measuring troponin I in blood samples, and the analysis may be performed close to the patient (near-patient test). The results may be obtained within 10 minutes, compared to approximately 1 hour for normal laboratory analysis.

The SpinChip Platform consists of the SpinChip hs-cTnI test and the SpinChip Analyzer and may be used at the emergency department to evaluate patients presenting with symptoms of acute myocardial infarction (chest pain). The test can use blood from finger prick or venous blood samples, either as whole blood or separated into plasma.

This multisite prospective, observational, non-randomized validation study will establish the repeatability, within-laboratory precision and reproducibility of the SpinChip hs-cTnI test for Li-Heparin plasma and repeatability for Li-Heparin whole blood samples. Potential subjects will be identified based on clinical cardiac troponin levels obtained from medical records at the individual study site.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Able and willing to provide signed written informed consent
  • Subjects >18 years old
  • Clinical cTn concentrations available

排除标准

  • Cognitive impairment precluding informed consent
  • Self-reported pregnancy
  • General clinical condition that affects the patient in a major way and the patient is considered to be in an unstable clinical condition
  • Previously included in the study with a sample within the required concentration ranges for the SpinChip hs-cTnI test
  • No venous blood sample collected for the study

结局指标

主要结局

Precision of the SpinChip hs-cTnI test using Li-heparin plasma and whole blood samples

时间窗: 1 day

Calculation of repeatability, within-laboratory precision and reproducibility of the SpinChip hs-cTnI test using cTnI results obtained from analysis of Li-Heparin plasma. Calculation of repeatability of the SpinChip hs-cTnI test using cTnI results obtained from analysis of Li-Heparin whole blood.

次要结局

  • Precision of the recommended commercially available control material(1 day)
  • Incidence of AEs, ADEs and DDs(1 day)