Assessment of Portable Oxygen Concentrators in Infants Undergoing Hypoxic Challenge Testing. A Randomized Crossover Study.

Hospital Son Espases1 个研究点 分布在 1 个国家目标入组 26 人开始时间: 2016年2月11日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
26
试验地点
1
主要终点
Assessment of portable oxygen concentrators (POCs) to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).

研究概览

简要总结

Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment. Onboard oxygen administration remains controversial. The Federal Aviation Administration approved portable oxygen concentrators (POCs) for onboard oxygen supply but there is lack of evidence about the use, especially in children. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

详细描述

According to a forecast by International Air Transport Association the number of people travelling on commercial aircrafts is predicted to rise up to 8.2 billion passengers in 2037. Therefore evidence-based flight recommendations will gain importance in the future, especially for patients suffering from chronic diseases. Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment where nitrogen is introduced in a plethysmograph reducing FiO2 (fraction of inspired oxygen) to 0.15. Oxygen supply is recommended if PaO2 (partial pressure of oxygen in arterial blood) drops <50-55 mmHg (in adults) or Sat.O2 (oxygen saturation) ≤ 85% (in infants) where non-invasive pulse oximetry is the recommended method for hypoxia assessment. Onboard oxygen administration remains controversial. The effectiveness of pulsed-dosed systems remains unclear in small patients due to shallow breathing patterns. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
None

入排标准

年龄范围
— 至 16 Years(Child)
性别
All
接受健康志愿者

入选标准

  • Patients with baseline Sat.O2 >94% AND
  • <1 year with neonatal respiratory disease
  • required oxygen supply in the last 6 months
  • chronic respiratory disease (eg cystic fibrosis, obstructive or restrictive pulmonary disease) with FEV 1 (forced expiratory volume in 1 second) or FVC (forced vital capacity) <50%.

排除标准

  • acute respiratory infection

研究组 & 干预措施

cPOC - pPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by cPOC (continuous-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by pPOC (pulsed-flow).

干预措施: continuous-flow portable oxygen concentrator (cPOC) (Device)

cPOC - pPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by cPOC (continuous-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by pPOC (pulsed-flow).

干预措施: pulsed-flow portable oxygen concentrator (pPOC) (Device)

pPOC - cPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by pPOC (pulsed-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by cPOC (continuous-flow).

干预措施: continuous-flow portable oxygen concentrator (cPOC) (Device)

pPOC - cPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by pPOC (pulsed-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by cPOC (continuous-flow).

干预措施: pulsed-flow portable oxygen concentrator (pPOC) (Device)

结局指标

主要结局

Assessment of portable oxygen concentrators (POCs) to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).

时间窗: 20 minutes

Hypoxia (Sat.O2 drop ≤ 85%) measured by Masimo SET Radical-7 Electron® pulse oximeter is induced performing HCT. Thereafter oxygen is administered through cPOC or pPOC until baseline Sat.O2 (Sat.O2 \>93%) is achieved.

次要结局

  • POC comparison to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)
  • Relate patient age (months) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)
  • Relate patient weight (kilograms) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)

研究者

发起方
Hospital Son Espases
申办方类型
Other
责任方
Principal Investigator
主要研究者

Sebastian Sailer

Principal Investigator

Hospital Son Espases

研究点 (1)

Loading locations...

相似试验