Respiratory Mechanics Assessment at Different Head of the Bed Elevations in Mechanically Ventilated Patients
试验速览
- 阶段
- 不适用
- 状态
- 招募中
- 入组人数
- 40
- 试验地点
- 2
- 主要终点
- Respiratory system compliance
研究概览
简要总结
The effects of different degrees of head-of-bed elevation on respiratory mechanics are poorly explored in the literature, and no study has investigated such effects using electrical impedance tomography, esophageal and gastric balloons to identify the ideal angle for optimizing respiratory mechanics. The hypothesis is that there is a optimal degree for the respiratory mechanics.
详细描述
Respiratory mechanics and regional ventilation will be monitored using electrical impedance tomography (Enlight 2100, Timpel Medical®, Brazil) . Esophageal and gastric pressures will be obtained through esophageal and gastric balloon catheters (Nutrivent®) (validation concerning to modified Baydur maneuver - slope delta esophageal pressure/delta airway pressure (0,8-1,2). We are using the hardware Pneumodrive (Biônica, Recife, Brazil) to record and store the esophageal, gastric and airway pressures, these data will be analyzed using LabVIEW 7.1 (Pneumobench).
Initially, patients will be positioned at 0 degrees of head-of-bed elevation, and after stabilization of the plethysmogram, data from electrical impedance tomography, hemodynamics, and arterial blood gas will be collected (arterial blood will be drawn by a nurse or physician). Sequentially and in the same manner, the bed will be adjusted to 10, 20, 30, and 40 degrees (the same data will be collected, except for the arterial blood sample, which will only be collected at the 40-degree elevation). Then, an alveolar recruitment maneuver will be performed, followed by a PEEP titration with 10-degree of head-of-bed elevation.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Basic Science
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients under invasive mechanical ventilation, intubated due to respiratory failure
排除标准
- •Hemodynamics instability, contraindication for monitoring with esophageal and gastric catheters, and Electrical impedance tomography, no authorization of medical team of the intensive care unit, and contraindication for lung recruitment maneuver
结局指标
主要结局
Respiratory system compliance
时间窗: At 0, 10, 20, 30, 40-degrees of head-of-bed elevation, and with titrated PEEP at 10-degrees of head-of-bed elevation
Respiratory system compliance (mL/cmH2O) will be measured using electrical impedance tomography monitoring (Enlight 2100, Timpel Medical®, Brazil).
Lung compliance
时间窗: At 0, 10, 20, 30, 40-degrees of head-of-bed elevation, and with titrated PEEP at 10-degrees of head-of-bed elevation
Lung compliance (mL/cmH2O) will be measured offline using the esophageal pressure tracings. By knowing the respiratory system and chest wall compliance, the lung compliance will be calculated. (1/respiratory system compliance = 1/chest wall compliance + 1/lung compliance)
Chest wall compliance
时间窗: At 0, 10, 20, 30, 40-degrees of head-of-bed elevation, and with titrated PEEP at 10-degrees of head-of-bed elevation
Chest wall compliance (mL/cmH2O) will be measured offline using the esophageal pressure tracings. Chest wall compliance = tidal volume / delta esophageal pressure
次要结局
- Hemodynamics satefy of keeping low degrees of head of the elevation(At 0, 10, 20, 30, 40-degrees of head-of-bed elevation)
- Gastric pressure(At 0 and 40-degrees of head-of-bed elevation, and with titrated PEEP at 10-degrees of head-of-bed elevation)
- Oxygenation(At 0 and 40-degrees of head-of-bed elevation, and with titrated PEEP at 10-degrees of head-of-bed elevation)
- Pressure between patient skin surface and the mattress(At 0, 10, 20, 30, 40-degrees of head-of-bed elevation)