The Influence of Standard Ventilator Setting Adjustments on Dyspnea Experienced in Awake Mechanically-ventilated Patients: A Pilot Study

Beth Israel Deaconess Medical Center1 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2026年9月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
尚未招募
入组人数
20
试验地点
1
主要终点
Change in dyspnea scale from baseline

研究概览

简要总结

In the past 5 years, there are increasing data suggesting that patients treated with mechanical ventilation experience shortness of breath, despite appropriate sedation. This adverse experience is believed to contribute to the finding that up to 25% of patients who survive severe respiratory diseases experience mental health problems including post traumatic distress syndrome (PTSD). The purpose of this study is to evaluate if/how sequential changes in the delivery of mechanical ventilation affect shortness of breath sensation in awake patients requiring mechanical ventilation. Improving the knowledge of the impact of the patient-ventilator interaction on shortness of breath sensation may lead to strategies to improve the comfort of non-sedated and sedated ventilated patients, and thereby reduce mental health sequelae in survivors of acute severe respiratory diseases The investigators hypothesize that current ventilator strategies, particularly reduced tidal volume (size of breath given by the ventilator) utilized in managing patients with severe respiratory diseases, contribute to shortness of breath in patients with increased drive to breathe. In this setting, some safe ventilator changes may improve or worsen the shortness of breath sensation in awake patients on mechanical ventilation.

详细描述

For patients who develop acute respiratory failure, endotracheal tube (ETT) intubation and mechanical ventilation represent a potentially life-saving intervention to provide support for breathing and allow the opportunity for lungs to recover from critical illness. Current standard practice for mechanical ventilation varies, and includes selecting mechanical ventilator mode (including volume-cycled or pressure-cycled), and routine adjustments of ventilator settings [including settings such as flow rate, tidal volume (Vt), positive end-expiratory pressure (PEEP)].

To assess and confirm the choice of mechanical ventilator settings for each patient, current standard practice includes monitoring vital signs, lung mechanics, observed use of accessory respiratory/breathing muscles by the patient, non-invasive measurement of percent oxygen saturation of the blood (SpO2), and direct blood sample analysis (including arterial or venous blood gas tests). However, direct assessment of patient breathing comfort, or the presence of breathing discomfort (i.e. dyspnea), is not routinely performed in mechanically ventilated patients, and represents a significant knowledge gap.

Dyspnea is a common symptom in hospitalized patients, but the incidence, prevalence and severity of dyspnea in mechanically ventilated ICU patients is incompletely understood. Dyspnea is a common symptom among hospitalized patients. It is defined by American Thoracic Society as "a term used to characterize a subjective experience of breathing discomfort that consists of qualitative distinct sensations that may vary in intensity. The experience derives from interaction among multiple physiological, psychological, social and environmental factors, and may induce secondary physiological and behavioral responses".

Limited publications suggest that up to 50% of intubated patients experience moderate-to-severe dyspnea during ICU admission (data limited to patients who are able to communicate), and survivors of critical illness acute respiratory failure report post-traumatic stress disorder (PTSD), and dyspnea symptoms other mental health problems commonly associated with including nightmares and flashbacks.

Dyspnea assessment in intubated and mechanically ventilated patients represents a substantial challenge. For patients with acute critical illness requiring intubation and mechanical ventilation, treatment often also requires a heavy significant level of analgosedation to provide patient comfort, but while rendering the patient non-communicative.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

盲法说明

Patients will be unaware of the nature of the changes in ventilator, the researcher obtaining the ratings and analyzing the data will be blinded.

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Admitted to a participating ICU at BIDMC
  • Requiring mechanical ventilation
  • Awake by both of the following criteria
  • RASS -2 to +2*
  • CAM-ICU negative‡
  • Able to communicate/answer dyspnea questionnaire
  • RASS (Richmond Agitation-Sedation Scale) is a validated scale commonly used in the ICU to assess patient's level of sedation and agitation. It ranges from -5 to +
  • While -5 means an unarousable coma, +4 means violent patient with immediate self danger. The scale from -2 to +2 ranges from light sedation to agitated.
  • CAM-ICU (Confusion Assessment Method) is an ICU validated scale used to assess the presence of delirium. A negative CAM-ICU scale has good positive and negative predictive value to diagnose or exclude delirium.
  • Physicians and nurses are trained to evaluate patients using both scales

排除标准

  • Age < 18 years old
  • Comfort measures only
  • Hemodynamic instability (MAP < 65 mmHg) or increasing requirement of vasopressor
  • PEEP > 10 cmH2O
  • FiO2 > 0.6
  • Current Prone position
  • Current Pneumothorax
  • Bronchopleural fistula
  • Neuromuscular conditions that impair responding to the dyspnea scale or expression
  • Prisoners
  • Pregnant women
  • Patients intubated and ventilated for less than 12 hours during the current ventilation episode
  • pH < 7.20 or > 7.55
  • Presence of chest tube
  • Status post thoracotomy
  • Treating clinician refusal

研究组 & 干预措施

Mechanical Ventilator Changes

Experimental

The intervention consists of changes in ventilator with evaluation of associated patient dyspnea.

The sequence of interventions will include:

  • After consenting and explaining the procedures in details, baseline dyspnea will be obtained

  • All patients will be set on volume control/assist control (VC/AC) with flow at ramp and appropriate settings to approximate their initial tidal volume, minute ventilation, respiratory rate, FiO2 and PEEP. This change is routinely performed in the ICU. Patients already on VC/AC will be kept on same settings. The investigators will check dyspnea after this initial change.

  • The following series of changes in mechanical ventilation will be instituted in a randomized sequence, lasting 3 minutes each:

  • Change in tidal volume

  • Change in inspiratory flow

  • Change in PEEP

  • Change to PSV to match their initial settings.

  • After every change the patient will rate dyspnea with an ordinal scale and Pocc and P01 will be assessed

干预措施: Change to PSV (Other)

Mechanical Ventilator Changes

Experimental

The intervention consists of changes in ventilator with evaluation of associated patient dyspnea.

The sequence of interventions will include:

  • After consenting and explaining the procedures in details, baseline dyspnea will be obtained

  • All patients will be set on volume control/assist control (VC/AC) with flow at ramp and appropriate settings to approximate their initial tidal volume, minute ventilation, respiratory rate, FiO2 and PEEP. This change is routinely performed in the ICU. Patients already on VC/AC will be kept on same settings. The investigators will check dyspnea after this initial change.

  • The following series of changes in mechanical ventilation will be instituted in a randomized sequence, lasting 3 minutes each:

  • Change in tidal volume

  • Change in inspiratory flow

  • Change in PEEP

  • Change to PSV to match their initial settings.

  • After every change the patient will rate dyspnea with an ordinal scale and Pocc and P01 will be assessed

干预措施: Change to VC/AC (Other)

Mechanical Ventilator Changes

Experimental

The intervention consists of changes in ventilator with evaluation of associated patient dyspnea.

The sequence of interventions will include:

  • After consenting and explaining the procedures in details, baseline dyspnea will be obtained

  • All patients will be set on volume control/assist control (VC/AC) with flow at ramp and appropriate settings to approximate their initial tidal volume, minute ventilation, respiratory rate, FiO2 and PEEP. This change is routinely performed in the ICU. Patients already on VC/AC will be kept on same settings. The investigators will check dyspnea after this initial change.

  • The following series of changes in mechanical ventilation will be instituted in a randomized sequence, lasting 3 minutes each:

  • Change in tidal volume

  • Change in inspiratory flow

  • Change in PEEP

  • Change to PSV to match their initial settings.

  • After every change the patient will rate dyspnea with an ordinal scale and Pocc and P01 will be assessed

干预措施: Change in Tidal Volume (Other)

Mechanical Ventilator Changes

Experimental

The intervention consists of changes in ventilator with evaluation of associated patient dyspnea.

The sequence of interventions will include:

  • After consenting and explaining the procedures in details, baseline dyspnea will be obtained

  • All patients will be set on volume control/assist control (VC/AC) with flow at ramp and appropriate settings to approximate their initial tidal volume, minute ventilation, respiratory rate, FiO2 and PEEP. This change is routinely performed in the ICU. Patients already on VC/AC will be kept on same settings. The investigators will check dyspnea after this initial change.

  • The following series of changes in mechanical ventilation will be instituted in a randomized sequence, lasting 3 minutes each:

  • Change in tidal volume

  • Change in inspiratory flow

  • Change in PEEP

  • Change to PSV to match their initial settings.

  • After every change the patient will rate dyspnea with an ordinal scale and Pocc and P01 will be assessed

干预措施: Change in inspiratory flow (Other)

Mechanical Ventilator Changes

Experimental

The intervention consists of changes in ventilator with evaluation of associated patient dyspnea.

The sequence of interventions will include:

  • After consenting and explaining the procedures in details, baseline dyspnea will be obtained

  • All patients will be set on volume control/assist control (VC/AC) with flow at ramp and appropriate settings to approximate their initial tidal volume, minute ventilation, respiratory rate, FiO2 and PEEP. This change is routinely performed in the ICU. Patients already on VC/AC will be kept on same settings. The investigators will check dyspnea after this initial change.

  • The following series of changes in mechanical ventilation will be instituted in a randomized sequence, lasting 3 minutes each:

  • Change in tidal volume

  • Change in inspiratory flow

  • Change in PEEP

  • Change to PSV to match their initial settings.

  • After every change the patient will rate dyspnea with an ordinal scale and Pocc and P01 will be assessed

干预措施: Change in PEEP (Other)

结局指标

主要结局

Change in dyspnea scale from baseline

时间窗: 3 minutes after the change

Change in dyspnea scale when changing the following settings Increase and decrease in tidal volume Increase, decrease in flow and change in flow curve Changes in PEEP Change to PSV to match their initial settings. We will use a modified multidimensional dyspnea scale, which includes from pleasant, neutral, slight sensation, annoying, distressing and unbearable. It also includes sensory qualities which ranges from 1 to 5 (5 being worse)

次要结局

  • Changes in dyspnea scales with change to VC/AC(3 minutes after change)
  • Change in respiratory drive measured by P0.1(3 minutes after each change)
  • Change in respiratory effort measured by Pocc(3 minutes after every change)
  • Changes in P0.1/Pocc ratio(3 minutes after every change)