Oxygen Hemoglobin Saturation in COVID Patients: Comparison of Arterial Gasometry and Pulse Oximetry

Assistance Publique - Hôpitaux de Paris2 个研究点 分布在 1 个国家开始时间: 2022年5月2日最近更新:
适应症

试验速览

阶段
不适用
状态
撤回
试验地点
2
主要终点
To evaluate the accuracy of SpO2 value provided by portable pulse oximeter in COVID patients compared with the reference method.

研究概览

简要总结

In many patients, respiratory Sars-Cov2 infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises. A large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation remains sometimes doubtful, or even absent from FDA standards.

The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method. The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.

详细描述

The constant monitoring of saturation by the pulse oximeter, the result of three centuries of technological progress, has introduced a true revolution in medical management that now relies on this vital parameter. In many patients, Sars-Cov2 respiratory infection causes arterial hypoxemia, which remains without signs of verbalized respiratory distress, up to a certain point. This phenomenon, called "happy" or "silent" hypoxemia, has a plural pathophysiological basis. Hypoxemia has been shown to be predictively associated with admission to the ICU. Therefore, the question of constant monitoring of oxygenation, practiced on a large scale, at home, in asymptomatic patients or contact cases, arises.

Despite its many advantages (non-invasive, kinetic), the pulse oximeter can be misused in several pathophysiological situations. The reference method for measuring hemoglobin oxygen saturation remains arterial gasometry.

However, a large number of portable pulse oximeter are currently freely available on the market; however, their clinical validation is sometimes questionable or even absent from FDA standards.

The objective of this study is to evaluate the accuracy of SpO2 values provided by portable pulse oximeter in COVID patients, in comparison with the reference method (arterial gasometry). The study will be conducted on a population of adult patients with COVID, hospitalized in the ICU, for whom gasometry sampling is already scheduled in the usual management.

Daily at 6am, during the systematic monitoring, arterial gasometry will be taken (as part of routine care) by the nurse or medical staff on duty. Synchronous measurement of Sp02 given by three portable pulse oximeter (1 measurement point per device, therefore 3 measurements/patient/day for a minimum of 3 days).

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients hospitalized in the intensive care unit for COVID
  • Age ≥ 18 years
  • Patients or family/trusted person informed and not objecting to participating in the research

排除标准

  • Carbon monoxide poisoning
  • Homozygous sickle cell disease
  • Severe anemia (< 7 g/L) with associated hypoxemia
  • Methemoglobinemia
  • Sulfhemoglobinemia
  • Sepsis, vasoconstrictive drugs
  • Patients under guardianship / curators
  • Patient under state medical aid

结局指标

主要结局

To evaluate the accuracy of SpO2 value provided by portable pulse oximeter in COVID patients compared with the reference method.

时间窗: once a day

Measurement of oxygen saturation by pulse oximetry once a day, at 6 a.m., for a minimum of 3 days, using three pulse oximeters (1 measurement point per device/day), in parallel with a synchronous, programmed arterial gasometry sampling Evaluation according to the standards of the SFBC (French Society of Clinical Biology) for the comparison of methods: minimum 40 points per apparatus, evenly distributed over the pathophysiological range of the saturations explored (9, 10): 60-100%.

次要结局

  • Identification of pathophysiological or analytical biases specific to COVID patients.(through study completion, an average of 7 months)