Electromagnetic Interference by Transcutaneous Neuromuscular Electrical Stimulation in Patients With Optimizer(R) - a Pilot Safety Study

Medical University of Vienna0 个研究点目标入组 6 人开始时间: 2015年5月最近更新:
适应症

试验速览

阶段
4 期
状态
已完成
入组人数
6
主要终点
The patients with CCM-system are assessed for EMI during TENS-application

研究概览

简要总结

Systemic skeletal muscle atrophy (cachexia) is a typical side effect of chronic heart failure.

Patients with cachexia can benefit from transcutaneous electrical neuromuscular stimulation (TENS). It is applied to strengthen the musculature and improve exercise tolerance and quality of life.

It is unclear if application of TENS can cause electromagnetic interference (EMI) in patients with Optimizer® pacemaker systems for improvement of heart failure.

This pilot safety study enrolled 6 patients with chronic heart failure and reduced left ventricular ejection fraction. All patients had an Optimizer® pacemaker system implanted for a minimum of 6 months, no patients had any ventricular arrhythmic episode 3 months prior to the study, and all were in clinically stable condition and willing to participate in this study.

Neuromuscular stimulation was administered via TENS in 4 different modes (HF, LF, Burst, MF) on the right and left musculus trapezius and on both musculi quadriceps femoris. All patients were tested for EMI without or with cardiac contractility modulation (CCM) delivery of the Optimizer®-system respectively.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Basic Science
盲法
None

入排标准

年龄范围
18 Years 至 99 Years(Adult, Older Adult)
性别
Male
接受健康志愿者

入选标准

  • >18 years, with implantation of the Optimizer ® >6 months ago, clinically stable condition and without any ventricular arrhythmic episode in the last 3 months.

排除标准

  • Local swelling or edema at the site of stimulation, intolerance to electrical stimulation, or presence of other metallic implants at the site of stimulation

结局指标

主要结局

The patients with CCM-system are assessed for EMI during TENS-application

时间窗: 6 months

Electromagnetic interference can cause artificial sensing in pacemaker or ICD system. These artefacts may have the effect of suppressing the pacemaker pulses sensing of tachycardia in ICD. The patients will be exposed to electrical stimulations, which will be of different forces on the neck and the upper thigh with r4x. These impulses and the effects will be noted down. It will be investigated if there is an interference with the optimizer ® depending on the different sites of the electrical stimulations and the different impulse forces. Many patients with an Optimizer ® have simultaneously a pacemaker - or ICD implant. A special extra monitor will note if there are probable interferences with these pacemaker- or ICD implants.