临床试验/NCT06623058
NCT06623058尚未招募不适用

Accupower ® HCV Performance Evaluation Quant Kit Bioneer Existation™FA 96/384

CerbaXpert1 个研究点 分布在 1 个国家目标入组 45 人开始时间: 2024年10月1日最近更新:
适应症

试验速览

阶段
不适用
状态
尚未招募
发起方
入组人数
45
试验地点
1
主要终点
Commutability between serum and plasma hepatitis C viral load results with the ACCUPOWER HCV Quant Kit Bioneer Existation™FA 96/384.

研究概览

简要总结

As part of the CE marking of a hepatitis C diagnostic kit, the South Korean manufacturer Bioneer wishes to set up a performance study in France in accordance with the IVDR (RE 2017/746).

Cerba Xpert CRO of the Cerba Healthcare group promoted this performance study by setting up a prospective collection of blood samples from patients suffering from the virus hepatitis C and whose viral load is positive. This prospective collection will be carried out in 4 laboratories of the Cerba Healthcare group.

详细描述

Viral hepatitis are primarily human systemic infections caused by viruses hepatic diseases which cause damage to the liver by hepatocyte infection of the virus and/or a host immune response to the virus. (1) Hepatitis is grouped into five types (A, B, C, D, E) and is mainly transmitted by parenteral, sexual and fetomaternal.(2) The hepatitis C virus (HCV) was discovered in 1989.(3) It is an RNA virus that belongs to the family Flaviviridae.(4) Infection with the hepatitis C virus is characterized by clinical signs such as such as jaundice, asthenia, anorexia. In approximately 30% (15% to 45%) of infected people, the virus is spontaneously eliminated within six months following infection, without any treatment. (5) For the remaining 70% (55% to 85%), the infection will progress to a chronic form. In patients chronic, the risk of cirrhosis is 15% to 30% in the 20 years following infection. (5) A assessment of liver damage and virus replication in the blood is made before starting treatment.(6) Detection/quantification of HCV RNA in serum or plasma must therefore be available before to initiate antiviral treatment. HCV RNA must be tested for by a sensitive test giving a result viral load expressed in IU/ml and Log IU/ml. (6) Monitoring of response to treatments is done also by the detection of viral RNA, particularly at 12 weeks post-treatment where a viral load undetectable is a sign of effective treatment and healing. (6) On the European market, quantitative determination of hepatitis C viral load as part of Diagnosis and monitoring of the disease is done by real-time PCR which is the reference technique.(7) As part of routine care, there are several diagnostic PCR kits that can be used on plasma or serum. Today, there are no CE approved PCR kits that allow the quantification of viral RNA at the both serum and plasma. Our study will make it possible to evaluate the performance of the Bionner kit PCR kit ACCUPOWER Quant Kit Bioneer Existation™FA 96/384 on serum and plasma of affected patients hepatitis C.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • The patient to be included in this study must be able to understand the purpose of the research, in order to to give free and informed consent.
  • Subject aged over 18
  • Subject presenting at the investigation center and meeting one of these two criteria:
  • Subject presenting with a prescription for HCV viral load determination
  • Subject whose HCV infection is previously confirmed by CE marked tests.
  • Subject capable of understanding the purpose of the research having given express free and informed consent
  • Subject affiliated to or beneficiary of a social security system

排除标准

  • Protected subject: adult under guardianship, curatorship or other legal protection, deprived of liberty by judicial decision or administrative Subject participating in a clinical study/receiving investigational treatment

结局指标

主要结局

Commutability between serum and plasma hepatitis C viral load results with the ACCUPOWER HCV Quant Kit Bioneer Existation™FA 96/384.

时间窗: 4 months

The submitted protocol is carried out as part of a 'commutability' study, that is to say for Compare two matrix types on the Accupower ® HCV Quant Kit Bioneer Existence™FA 96/384. Indeed, for these two matrices to be claimed in the IFU of the kit it is necessary to prove that the two matrices give the same result. The planned size of the evaluable cohort is 45 patients included over a period of 4 month. This size is justified by the recommendations of the applicable CLSI EP09 guide. This guide details the use of several mathematical models that guarantee robust analysis even on a small cohort size. The operating methodology of the results is detailed below, taken from the recommendations of the CLSI EP09 guide

次要结局

未报告次要终点

研究者

发起方
CerbaXpert
申办方类型
Other
责任方
Sponsor

研究点 (1)