VerioVue Enhancements Regulatory Clinical Evaluation - Professional Comparison to Validated Method: Neonate Study
试验速览
- 阶段
- 不适用
- 状态
- 终止
- 发起方
- 入组人数
- 99
- 试验地点
- 3
- 主要终点
- Performance evaluation: blood glucose monitoring system accuracy: BGMS vs reference instrument
研究概览
简要总结
The goal of this performance evaluation is to compare the blood glucose results obtained using the VerioVue Blood Glucose Monitoring system (BGMS) with those obtained from a validated comparator method (iSTAT 1 Analyser) using blood obtained from neonates. The main question it aims to answer is:
•How accurate is the VerioVue BGMS when compared to a product that has already been confirmed as accurate (iSTAT 1 Analyser) when hospital staff test blood taken from neonates on these instruments? Participants (neonates) will have a small amount of blood taken from a heel prick (performed by a HCP for medical purposes) or from an existing arterial line.
详细描述
UK NHS Hospital sites (between 1 and 3 sites) will be used to collect a minimum of 200 evaluable blood samples from neonates meeting the inclusion criteria. Consent will be taken from the neonates parent prior to participation in the study. They will consent to one or two blood samples to be taken from their child. If the participant meets the enrolment criteria then demographics and prescription medication (including dietary supplements) information will be collected. A small volume of blood (max 150µl) will be collected from an existing arterial line or from a heel prick that is being performed for medical purposes. This blood will then be applied to six investigational One Touch VerioVue meters (with expanded haematocrit range of 20-65%) fitted with OneTouch Verio test strips and the blood glucose meter readings recorded. Three different lots of test strips will used in rotation across the study. Within 5 minutes of the last meter test, NHS study staff will use blood from the same sample and perform a test on the iSTAT 1 Analyser to obtain blood glucose, haematocrit and oxygen levels of the sample. No treatment decisions will be made based on the blood glucose results obtained for this study. The study documentation will be reviewed by LifeScan monitors for completeness and accuracy and the results entered into a validated database. The data will be extracted and statistically analysed to determine the accuracy of the blood glucose results obtained using the OneTouch VerioVue BGMS when compared to the blood glucose results obtained on the iSTAT 1 Analyser.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- 0 Days 至 28 Days(Child)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Age - Newborn babies, 28 days old or less with a corrected gestational age of at least 34 weeks and a weight of at least 1700g or more at the time of participation.
- •Informed Consent - Study participants parents must read the Parental Participant Information Sheet and sign the Parental Informed Consent Form
- •Study participants parent agrees to provide information related to demographics, prescription medication, dietary supplements, and results from other physician ordered testing e.g., Bilirubin (total, conjugated, and unconjugated) where available.
- •Study participants parent agrees to allow study staff access to medical records where necessary.
- •Study participants parents agree to all aspects of the study process, including where applicable arterial blood draw from an existing arterial line or a heel prick performed by an HCP for medical purposes.
排除标准
- •Age- Newborn more than 28 days old.
- •Study participants with a gestational age of less than 34 weeks at the time of participation.
- •Study participants with a gestational weight of less than 1700g
- •Current positive test result for Covid-19.
结局指标
主要结局
Performance evaluation: blood glucose monitoring system accuracy: BGMS vs reference instrument
时间窗: 1 hour
Performance Evaluation to compare the VerioVue BGMS blood glucose results to results obtained using the iSTAT 1 Analyser, a validated comparator method.
次要结局
未报告次要终点