Comparative Study of the Topcon Tonometers CT-800 and TRK-2P and the Haag-Streit Goldmann Manual Tonometer (Predicate) to Demonstrate Conformance to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification [510(k)] Submissions, and to the Applicable Supplemental Information Sheet and Comparison of Pachymetry Values for Topcon TRK-2P With the Topcon SP-1P (Predicate)
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 345
- 试验地点
- 1
- 主要终点
- Agreement of IOP
研究概览
简要总结
The study will assess the CT-800 and TRK-2P for conformity to ANSI Z80.10-2014 Ophthalmic Instruments - Tonometers, to the FDA Guidance for Industry and FDA Staff, Tonometer-Premarket Notification [510(k)] Submissions, and to the applicable Supplemental Information Sheet. This clinical device study will also assess the substantial equivalence of TRK-2P to a predicate device in regard to pachymetry.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •be at least 18 years of age of either sex and any race or ethnicity;
- •be willing and able to provide written informed consent prior to any study procedures being performed;
- •be willing and able to follow all instructions and attend all study visits;
排除标准
- •have only one functional eye;
- •have poor or eccentric fixation in either eye;
- •have corneal scarring or have had corneal surgery, including corneal laser surgery in either eye;
- •have microphthalmos in either eye;
- •have buphthalmos in either eye;
- •be a contact lens wearer, meaning having worn soft contact lenses within the past 3 months and/or rigid permeable gas lenses within the past 6 months;
- •have dry eyes, meaning having been diagnosed by a physician with dry eyes and currently using a prescribed medication or daily use of an artificial tear;
- •be a lid squeezer - blepharospasm;
- •have nystagmus in either eye;
- •have keratoconus in either eye;
- •have any other corneal or conjunctival pathology or infection in either eye;
- •have a condition or a situation, which in the Investigator's opinion, may put the subject at increased risk, confound study data, or interfere significantly with the subject's study participation.
结局指标
主要结局
Agreement of IOP
时间窗: 1 day
Agreement of the measured IOP between the test devices and the predicate device for the tonometer function of CT-800 and TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and linear regression.
Agreement of Corneal Thickness
时间窗: 1 day
Agreement of measured corneal thickness between the test device and the predicate device for the pachymeter function of the TRK-2P. Presentation of results and analyses will include descriptive statistics (mean, standard deviation, limits of agreement), Bland-Altman plot and Deming regression.
次要结局