Clinical Accuracy Validation of the Medical Body Composition Analyzer Seca mBCA 555 in Comparison to Seca mBCA 515

Seca GmbH & Co. Kg.1 个研究点 分布在 1 个国家目标入组 120 人开始时间: 2019年6月20日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
120
试验地点
1
主要终点
SMM

研究概览

简要总结

The mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations (BCA-01, BCA-02, BCA-20) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated based on the previous study BCA-03 conducted at the UKE.

The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.

In order to apply prediction equations and normal ranges from previous clinical investigations BCA-01, BCA-02, BCA-03 and BCA-20 to the new mBCA 555 it shall be compared to the mBCA 515.

详细描述

The use of bioelectrical impedance analysis (BIA) is widespread both in healthy subjects and patients. BIA allows the determination of the fat-free mass (FFM), skeletal muscle mass (SMM), fat mass (FM), extracellular water (ECW) and total body water (TBW) when using appropriate population, age or pathology-specific BIA equations and established procedures. BIA equations are based on validation studies where BIA is compared with reference methods such as Air Displacement Plethysmography (ADP), Dual-Energy X-Ray Absorptiometry (DXA), Magnetic Resonance Imaging (MRI) and isotope dilution.

The predicate device seca mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations in Kiel (Germany) and New York City (USA) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated in a previous study conducted at the University Hospital Hamburg-Eppendorf.

The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.

Primary objective is to test whether the mBCA 515 prediction equations can be applied to the new mBCA 555. If necessary, to proof substantial equivalence, prediction equations should be adjusted for the mBCA 555.

Secondary objective is to compare Impedance, phase angle, reactance and resistance for different frequencies in order to clarify if normal ranges for BIVA and phase angle generated with the mBCA 515 need to be adjusted for the new mBCA 555.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • male and female subjects,
  • 18 years and older,
  • subjects need to be able to sign the informed consent and privacy policy

排除标准

  • acute disease
  • pregnancy
  • intake of diuretics
  • edema diagnosed by inspection and palpation of lower limbs
  • paralysis e.g. after a stroke
  • neurodegenerative diseases e.g. ALS
  • tumors in treatment
  • amputation
  • electronic implants e.g. pacemaker
  • probands who cannot provide an ICF by themselves
  • probands who might be dependent from the sponsor or the investigation site
  • current alcohol abuse
  • active prostheses
  • electronic life-support systems, e.g. artificial heart, artificial lung
  • portable electronic medical devices, e.g. ECG devices or infusion pumps
  • metallic implants

结局指标

主要结局

SMM

时间窗: 10min

Skeletal muscle mass, bias of mBCA 555 in comparison to mBCA 515

次要结局

  • Phase Angle(10 minutes)