临床试验/JPRN-UMIN000044712
JPRN-UMIN000044712已完成未知

Accuracy of non-invasive cardiac output measurement in awake patients - Accuracy of non-invasive cardiac output measurement in awake patients

agoya University0 个研究点目标入组 40 人开始时间: 2021年9月1日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
40

研究概览

简要总结

The paired esCCO and TDCO measurements were compared. The respective bias (standard deviation) values were 0.13(0.60) L/min before extubation, and -0.48(0.78) L/min after extubation. The percentage errors were 25.1% before extubation and 29.6% after extubation. Thus, the accuracy of the noninvasive esCCO measurement is clinically acceptable as an alternative to TDCO measurement under different respiratory conditions of mechanical ventilation and spontaneous respiration.

研究设计

研究类型
Observational

入排标准

年龄范围
20years-old 至 ot applicable(—)
性别
All

入选标准

  • 未提供

排除标准

  • 1. Emergency surgery