跳至主要内容
临床试验/NCT03595501
NCT03595501已完成不适用

Analytical and Clinical Performance Testing Plan

Sight Diagnostics1 个研究点 分布在 1 个国家目标入组 679 人开始时间: 2018年6月20日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
679
试验地点
1
主要终点
Repeatability of CBC parameters provided by the OLO device

研究概览

简要总结

Clinical and analytical tests will be performed based on risk assessment and system specifications to verify that the performance of the investigational device is in accordance with its specifications.

详细描述

Comparison and precision studies will be conducted in up to 5 clinical sites in Israel and the US, and include:

  • Precision
  • Sample Matrix Comparison
  • Method comparison study and flagging analysis
  • Reference interval range

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
3 Months 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • For Residual Samples:
  • Specimen obtained by venipuncture or finger prick and collected into tubes normally used by the site
  • Patient is at least 3 months of age
  • Samples within 8 hours from phlebotomy
  • For Prospectively collected samples:
  • Subject is at least 22 years of age
  • Non-diseased individuals or, for specific studies, individuals with blood count ranges to cover indicated medical decision points and ranges
  • Samples within 8 hours from phlebotomy

排除标准

  • Exclusion criteria post blood draw and pre sample scan - For whole blood samples:
  • Visibly hemolyzed or clotted specimens
  • Specimens with insufficient blood volume to complete the procedure
  • Samples older than eight hours
  • Exclusion criteria post sample scan:
  • Instrument failure or sample rejected by the instrument due to system error or sample mishandling
  • The daily quality control sample measurements indicate that the assay run is outside the specifications for the instrument
  • Operator related error documented in the study records
  • Failure to adhere to study specifics or protocols

结局指标

主要结局

Repeatability of CBC parameters provided by the OLO device

时间窗: 3 months

Performance of the OLO device will be measured through repeatability of 20 replicates for 11 residual samples on each site. Samples collected for this study will include 4 within lab reference range, 3 around lower medical decision levels for HGB, PLT and WBC, and 4 around the upper range for RBC, HGB, WBC and PLT to cover all pathological levels and medical decision points. Standard deviation (SD) and coefficient of variation (CV) will be computed for each measurand per sample by site.

Sample Matrix Comparison

时间窗: 3 months

Paird capillary and venous whole blood samples will be collected. The samples will be analyzed in duplicate on the Sight OLO device. Analysis will include Passing-Bablok Regression analysis per parameter (Bland Altman plots, slope, intercept, with 95% confidence intervals, correlation coefficient, and % bias), between: The average of venous whole blood samples scans and average of capillary whole blood sample scans from the same individual on the Sight OLO device

Reproducibility of CBC parameters provided by the OLO device

时间窗: 3 months

Reproducibility studies will be conducted using 3 levels of commercial control materials (low, normal and high) to measure all CBC reported parameters. Control material will be run on two instruments at each site. Standard deviation (SD) and coefficient of variation (CV) will be calculated for each measurand for: (1) Between lots/sites; (2) Between instruments; (3) Between days; (4)Between operators/runs; (5) within-run and (6) total variability

Method Comparison

时间窗: 3 months

Evaluate the performance of the OLO device in comparison to values achieved with the predicate. Analysis will include Regression parameters (slope, intercept with 95% CI) between measurement of Sight OLO and measurement on predicate (Correlation coefficient, Bland Altman plots and overall % bias between predicate and Sight OLO device including % bias at medical decision points).

Reference Interval Range

时间窗: 3 months

Establish adult venous and fingerprick reference intervals for the OLO device. The non-parametric method will be used to calculate the lower and upper limits of the reference range.

次要结局

未报告次要终点

研究者

发起方
Sight Diagnostics
申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验