Validation of the Blood Pressure Monitoring Function of a Fingertip Photoplethysmography

RadiRad Co., Ltd.0 个研究点目标入组 170 人开始时间: 2023年10月20日最近更新:
适应症

试验速览

阶段
不适用
状态
尚未招募
入组人数
170
主要终点
Difference in blood pressure (mean)

研究概览

简要总结

High blood pressure is a major risk factor for cardiovascular diseases, stroke, and kidney disorders. Accurate blood pressure monitoring is crucial for the diagnosis, treatment, and prevention of complications related to high blood pressure. In recent years, due to the rapid development of wearable devices and mobile technology, wearable blood pressure monitors have gradually become a non-invasive and convenient method for blood pressure monitoring. However, the accuracy of these devices has not been fully established. This study aims to validate the performance of RadiHeart (an application program) in blood pressure measurement.

详细描述

Following the ISO 81060-2:2018 standard, the accuracy of RadiHeart will be evaluated by comparing the blood pressure values obtained by the device with those measured by a conventional mercury sphygmomanometer. This research is expected to recruit 170 adults aged 20 and above. All participants will undergo blood pressure measurements using both RadiHeart and a mercury sphygmomanometer. Participants will operate RadiHeart under the guidance of the testing personnel, while the mercury sphygmomanometer will be operated by trained healthcare professionals. The blood pressure measurement results from both devices will be recorded simultaneously, and the consistency between the blood pressure values measured by RadiHeart and the traditional mercury sphygmomanometer will be compared to validate the accuracy of the blood pressure application.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
20 Years 至 99 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • The subjects should be aged 20 or older, with an equal gender distribution of 50% male and 50% female. From this group, 85 subjects with reasonable data values will be selected, with a requirement that at least 30% of them are male and at least 30% are female. Additionally, they must meet the following blood pressure conditions:
  • At least 5% (5 individuals) of the participants should have reference systolic blood pressure readings ≥ 160 mmHg (21.33 kPa).
  • At least 20% (17 individuals) of the participants should have reference systolic blood pressure readings ≥ 140 mmHg (18.66 kPa).
  • At least 5% (5 individuals) of the participants should have reference systolic blood pressure readings ≤ 100 mmHg (13.33 kPa).
  • At least 5% (5 individuals) of the participants should have reference diastolic blood pressure readings ≤ 60 mmHg (8.0 kPa).
  • At least 20% (17 individuals) of the participants should have reference diastolic blood pressure readings ≥ 85 mmHg (11.33 kPa).
  • At least 5% (5 individuals) of the participants should have reference diastolic blood pressure readings ≥ 100 mmHg (13.33 kPa).

排除标准

  • Having any of the following conditions:
  • Arrhythmia
  • Medical assessment indicates that poor blood circulation would affect the blood pressure data collection.
  • Involuntary hand movements that would affect the blood pressure data collector.
  • Nail polish on the light sensor area of the fingertip oximeter.
  • Other conditions as determined by a physician that make the individual unsuitable for participation in the trial.

结局指标

主要结局

Difference in blood pressure (mean)

时间窗: 1 hour per individual

The mean value of the differences in blood pressure measured by RadiHeart and the reference sphygmomanometer should be within or equal to ±5.0 mmHg.

Difference in blood pressure (standard deviation)

时间窗: 1 hour per individual

The standard deviation of the differences in blood pressure measured by RadiHeart and the reference sphygmomanometer should be no greater than 8.0 mmHg.