临床试验/NCT04750642
NCT04750642已完成不适用

Cochlear Implant With Dexamethasone Eluting Electrode Array (The CI-DEX Study): Pivotal Study

Cochlear16 个研究点 分布在 3 个国家目标入组 42 人开始时间: 2021年10月4日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
Cochlear
入组人数
42
试验地点
16
主要终点
Comparison of Monopolar (MP1+2) Impedance Measurement (kOhms) Between Randomization Groups at 6 Months Postoperative

研究概览

简要总结

An evaluation of Cochlear's cochlear implant electrode array which passively elutes dexamethasone for a defined period of time to help reduce inflammatory responses.

详细描述

The primary objectives of the pivotal evaluation is to show the efficacy of the dexamethasone eluting electrode through reduction in electrode impedances as compared to a standard electrode and the improvement of speech recognition from preoperative baseline. The secondary objectives are to assess the benefit-risk balance of a dexamethasone eluting electrode as similar to a standard electrode by comparison of adverse events and speech outcomes.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Double (Participant, Care Provider)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Post-lingual, bilateral, moderate (≥ 40 dB HL) to profound sensorineural hearing loss at 250, 500and 1000 Hz and profound high-frequency hearing loss, defined by a pure-tone average (PTA) threshold, 2000 through 8000 Hz, ≥ 90 dB HL.
  • 18 years or older at time of consent.

排除标准

  • Abnormal cochlear and middle ear anatomy
  • History with cochlear implant surgery
  • Allergy to dexamethasone
  • Women who are pregnant or plan to become pregnant
  • Unable/unwilling to comply to study requirements

结局指标

主要结局

Comparison of Monopolar (MP1+2) Impedance Measurement (kOhms) Between Randomization Groups at 6 Months Postoperative

时间窗: Six months postoperative

The impedance was measured across the active electrodes of the cochlear implant for each participant. Reduced impedances indicate less fibrotic obstruction caused by trauma associated with the electrode insertion.

Change in Speech Perception Performance (Pre-recorded CNC Words in Quiet in Sound Booth Testing) Compared to Pre-operative Baseline Measured at 6 Months Postoperative With CI632D

时间窗: Six months postoperative

Scores on the CNC Words in Quiet test were expressed as a percentage of words repeated correctly by the participant in a sound booth. A higher score indicates better speech perception performance. This outcome measure included the CI632D group only.

次要结局

  • Comparison of Rate and Type of Device Related Adverse Events Between Randomization Groups(Twelve months postoperative)
  • Comparison of Speech Perception Performance (Pre-recorded CNC Words in Quiet in Sound Booth Testing) Between the Randomization Groups at Six Months Postoperative(Six months postoperative)
  • Comparison of Speech Perception Performance (AzBio Sentences in Quiet in Sound Booth Testing) Between Randomization Groups at Six Months Postoperative(Six months postoperative)