Validation of VoluSense Pediatrics, a New Method of Infant Spirometry Based on Electromagnetic Inductance Plethysmography
试验速览
- 阶段
- 不适用
- 发起方
- 入组人数
- 50
- 试验地点
- 1
- 主要终点
- Tidal volume. Unit is milliliter
研究概览
简要总结
This is a cross-sectional observational study, aiming to validate the accuracy of tidal breathing measurements in infants made by VoluSense Pediatrics, a lung function method based on electromagnetic inductance plethysmography, compared to an ultrasonic flowmeter. VoluSense Pediatrics consists of a flexible vest placed around the torso of the baby, and changes in tidal volume and flow is measured. The study will enable a better understanding of the usefulness of this equipment. The equipment is made by VoluSense Norway AS, and is owned by Haukeland University Hospital.
详细描述
All infants will be assessed at the pediatric department at Haukeland University Hospital, and all data obtained by the same two investigators (Bentsen and Olsen). The infants will be dressed in the appropriate-sized vest selected according to the length from the armpit to the hip of the child. Care will be taken to ensure that the vest fits snugly around the torso and that the width of the copper wire zig-zag coveres the entire thorax and abdomen including the pubic region. The vest will be applied directly to the skin; neither body nor diaper will be worn underneath. The infants will then be fed before they will be placed supine in a cot and encouraged to sleep. No sedation will be used. Once the infant has fallen asleep, data will be collected with VSP for 2-3 minutes before the Exhalyzer D® neonatal facemask connected in series with the ultrasonic flow sensor will be placed gently, but firmly over the infant's mouth and nose, ensuring a good seal. Once the facemask is in place, a marker will be inserted in the VSP recording. The Exhalyzer D® measurement will be started about thirty seconds after the facemask has been put on, to allow adaptation to the facemask. Simultaneous recording using both devices will then be done for about 1 minute before the mask will be removed. New markers will be inserted in the VSP recording as the Exhalyzer D® measurement is started and stopped facilitating analysis of the same breathing cycles.
Data will be analysed and graphs will be created with SPSS version 22 (IBM SPSS Statistics, New York, US) and MedCalc version 13.1 (MedCalc Software, Mariakerke, Belgium). Bland-Altman plots will be used to visualize agreement between the two measurement techniques. For each pair of measurements, the pairwise mean of the breathing parameters will be plotted on the x-axis, and the pairwise difference expressed as a percentage of their mean value on the y-axis. The mean difference and the 95% limits of agreement and their confidence intervals will be represented by horizontal lines in the graphs.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 2 Days 至 4 Weeks(Child)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Healthy term born or preterm born babies ≥ 2000 grams.
排除标准
- •Being in an unstable clinical condition.
- •Weight < 2000 grams.
结局指标
主要结局
Tidal volume. Unit is milliliter
时间窗: 1 hour
Tidal volume will be measured using the two Methods and compared using Bland-Altman statistics
次要结局
- Respiratory Rate. Unit is counts per minute(1 hour)
- Peak tidal expiratory flow (TPTEF). Unit is ml/sec(1 hour)
- the ratio of inspiratory to expiratory time (Ti/Te). This is a ratio, and thus no unit(1 hour)
- The ratio of tidal expiratory flow at 50% of expired volume to peak tidal expiratory flow (TEF 50/PTEF).(1 hour)
- Total expiratory time (Tptef/Te). This is a ratio, and thus no unit(1 hour)