临床试验/NCT01699763
NCT01699763已完成不适用

Evaluation of Blood Glucose Meter Systems - Contour® NEXT LINK Study

Ascensia Diabetes Care2 个研究点 分布在 1 个国家目标入组 112 人开始时间: 2012年10月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
112
试验地点
2
主要终点
MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range

研究概览

简要总结

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

详细描述

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and two additional Blood Glucose Monitoring Systems (BGMS) from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the three systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing was performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Males and females, 18 years of age or older
  • Willing to complete all study procedures

排除标准

  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

结局指标

主要结局

MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range

时间窗: 8 hours

Using the overall Blood Glucose (BG) range (34 to 561 mg/dL according to YSI subject plasma results), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI plasma results (BG reference) were compared. MARD is calculated from the sum of all \|(BG meter)-(BG reference)\|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 3 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value. Higher MARD value indicates larger difference between meter value and the reference value.

次要结局

  • MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)(8 hours)
  • MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range (<=80 mg/dL)(8 hours)

研究者

发起方
Ascensia Diabetes Care
申办方类型
Industry
责任方
Sponsor

研究点 (2)

Loading locations...

相似试验

Evaluation of Blood Glucose Monitoring Systems | 临床试验