Assessment of Ventilation Using Respiratory Volume Monitor During Sedation With Intravenous Propofol as Compared to Capnography in Patients Receiving Intravenous Sedation.
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 95
- 试验地点
- 1
- 主要终点
- Change in minute ventilation, tidal volume, and end-tidal carbon dioxide following propofol
研究概览
简要总结
A. Determine if Respiratory Volume Monitor (RVM) can be used during procedural sedation to adequately monitor patient's respiratory status as compared to capnograph Aims B. Compare the time of capture of respiratory events (depression, arrest, etc...) on both the RVM and the capnograph C. Compare the incidence of missing or poor quality readings between the RVM and the capnograph D. Compare RR readings between the RVM and the capnograph (during periods with adequate data quality on both devices) E. If RMV can identify respiratory obstruction. Capnography
详细描述
Rationale for Proposed Research: Continuous respiratory monitoring using the ExSpiron System will be particularly beneficial to patients receiving intravenous propofol breathing spontaneously with natural airway. Assessment and management of respiratory function and early intervention when indicated is a multifaceted, complex task often complicated by the unavailability of a cohesive and continuous monitoring system to guide clinical decisions. The ExSpiron system is designed for these patients and is intended to address some of the limitations of the current generation of hospital monitors and to provide healthcare providers with continuous real-time data regarding the patient's respiratory status. Recent RVM data has shown that stratification of patients based on the MV % of predicted, prior to opioid dosing has made it possible to identify patients who are at risk for further decrease in MV and opioid-induced respiratory depression (OIRD).
- In the spontaneously breathing, non-ventilated patient, current monitoring devices do not provide continuous, objective non-invasive, continuous real time information of important respiratory parameters such as TV, MV and RR.
- Current monitoring of non-intubated patients mostly relies on oximetry data, subjective clinical assessment and occasionally EtCO2 measurements. In appropriate clinical settings, the ExSpiron system can provide direct quantitative measure of ventilation parameters for use in clinical assessment and trend prediction as well as response to medications and other interventions.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- 20 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Inclusion Criteria: patient's ≥ 20 years of age receiving intravenous sedation, or intravenous general anesthesia with propofol.
排除标准
- •Criteria: patients not able to give informed consent, Inability to place the PadSets on the thorax. Existence of an active implanted neurostimulator.
结局指标
主要结局
Change in minute ventilation, tidal volume, and end-tidal carbon dioxide following propofol
时间窗: Starting approximately 10 min prior to colonoscopy and continuing until end of colonoscopy (average time: 45-60 minutes)
Minute ventilation (MV) and tidal volume (TV) measured by the Respiratory Volume Monitor will be reported as percent of baseline established during normal breathing prior to sedation. The percent decrease in MV and TV following propofol will be compared to changes in end-tidal CO2 measured by capnography. The amount of time between propofol doses and changes in MV and TV and EtCO2 will be quantified and compared.
次要结局
- Comparison of Respiratory Rate measurement by Respiratory Volume Monitor and Capnography(Starting approximately 10 min prior to colonoscopy and continuing until end of colonoscopy (average time: 45-60 minutes))
研究者
Bhavani Shankar Kodali
Staff Anesthesiologist
Brigham and Women's Hospital
