跳至主要内容
临床试验/NCT05790616
NCT05790616招募中不适用

Accuracy of Pulse Oximeters During Profound Hypoxia

Philips Clinical & Medical Affairs Global1 个研究点 分布在 1 个国家目标入组 66 人开始时间: 2023年1月28日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
入组人数
66
试验地点
1
主要终点
Algorithm validation

研究概览

简要总结

This study is designed to calibrate and determine the accuracy for SpO2, pulse rate and respiratory rate of the newly in-house build Test Device wrist 1 (TDw1, or EVA) at Philips.

SpO2, pulse rate and respiratory rate during hypoxia will be calculated by using data of well-known reference devices, including:

  • A reference SpO2 sensor of Nellcor placed at a fingertip, that reflects also continuously the pulse rate Will be used to compare with the test device.
  • A reference respiratory rate device that calculates the respiratory rate based on detection of end-tidal CO2 peaks by capnography.
  • Oxygen saturation in arterial blood samples (SaO2), determined by a co-oximeter will be used to calculate the accuracy of the test device.

During the study the following devices will be additionally used by the volunteers:

  • AppleWatch 7
  • TDw2, watch build by philips, using the PPG and software technology developed by Philips
  • A smartphone that detects reflected PPG signals from the reflected screen at the handpalm, by the build in frontfacing camera (TDc) of the smartphone

Volunteers will undergo progressive hypoxia (9 min/% O2) from 21 to 10% O2 in an altitude room, resulting in a volunteer's SpO2 of 73%. During this deliberated hypoxia, the volunteers wear the test and reference devices.

This study consists of 4 sub-studies (NI = non-invasive; IN = invasive with an arterial line):

  • NI (Fast-Sitting): volunteers are seated in the hypoxia room in which the ambient oxygen concentration decreases at a speed of 9 min/% O2. If the volunteer reached a SpO2 ≤73% for more than 1 minute, he/she leaves the hypoxia room. And will breath air with 21% oxygen. Volunteers wear TDw1 and TDw2 and the reference devices.
  • NI(Fast-Lying): identical to NI(Fast-Sitting) but volunteers lay on a mattress. Volunteers wear TDw1 and AppleWatch 7 and the reference devices.
  • NI (Slow-Sitting): identical to NI (Fast-Sitting), but after one of the volunteers reaches a SpO2 ≤73% for more than one minute, oxygen in the room increases at a speed of 9 min/% O2 until normal ambient air oxygen concentration of 21%. Volunteers wear TDw1 and AppleWatch 7 and the reference devices.
  • IN(Fast-Sitting): identical to NI(Fast-Sitting) but the volunteer's oxygen saturation in blood samples withdrawn via an arterial line is measured in the laboratory.

The NI studies include 18 healthy participants in each sub-study. After the first volunteers have completed the study, small adaptations in the software of the study devices is still possible, e.g. to increase the quality of the PPG-signals. After the three sub-studies are completed, the algorithm for conversion of raw PPG signals to SpO2, pulse rate and respiratory rate will be defined and fixed for the test devices.

During the IN-study, which can only be started after completion of all NI studies, an arterial catheter will be inserted in the radial artery of the 12 participating volunteers, in order to take several blood samples to measure oxygen saturation in the blood (25 samples at well-defined moments during the study per volunteer). Using these results of arterial oxygen saturation, the accuracy of the test devices can be calculated.

研究设计

研究类型
Interventional
分配方式
Non Randomized
干预模型
Single Group
主要目的
Screening
盲法
None

入排标准

年龄范围
18 Years 至 65 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Healthy subjects (ASA 1 and ASA 2) between 18 and 65 years of age upon signing the informed consent. (ASA: American Society of Anesthesiologist's health score). In each group, at least 30% of one gender need to be present, and at least 15% of participants have a Fitzpatrick's skin color scale of 5 or 6.

排除标准

  • Heavy smokers or individuals exposed to high levels of carbon monoxide that result in elevated carboxyhemoglobin levels.
  • Individuals subject to conditions that result in elevated levels of methemoglobin.
  • Individuals with low SpO2 or SaO2 (<95%) at FiO2 = 0,21
  • Moderate to severe claustrophobia for small rooms
  • Severe obesity (BMI > 35 kg/m2)
  • Hemoglobinopathies
  • Arteriopathies, including Raynaud disease
  • ASA 3 to 5 subjects
  • Tattoo's, skin lesions,use of skin lotions or use of nail polish/fake nails at the region of interest for TDw and TDc
  • Pregnancy

结局指标

主要结局

Algorithm validation

时间窗: 90 minutes per test per participant

1. the calculation and validation of the algorithm for 1. conversion of the modulation ratio R of theraw PPG signals to SpO2 for all wrist Test Devices. 2. conversion of PPG signals to PR and RR. 2. determining the accuracy of TDw1 and TDc with cooximetry as golden benchmark in the IN study. 3. Looking for the feasibility of TDw2 for further development The accuracy is represented by A(rms), and specifically for the NI study, the SpO2 range is continuous (SpO2 100- 73%, about 16 000 samples per volunteer). During the IN study the accuracy is represented by A(rms) and calculated for SpO2 versus Sa O2 (about 20 to 25 samples per volunteer).

次要结局

  • Plateau 3(90 minutes per test per participant)
  • Statistics pooled subjects(90 minutes per test per participant)
  • Statistics individual subject 2(90 minutes per test per participant)
  • Plateau 1(90 minutes per test per participant)
  • Plateau 2(90 minutes per test per participant)
  • Plateau 4(90 minutes per test per participant)
  • Statistics individual subject(90 minutes per test per participant)

研究者

申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验