Comparative Analysis of the Entegrion Point of Care Coagulation Monitor (PCM) With ROTEM in the Measurement of Viscoelastic Coagulation Parameters

Entegrion, Inc.3 个研究点 分布在 1 个国家目标入组 120 人开始时间: 2018年8月20日最近更新:
适应症

试验速览

阶段
不适用
入组人数
120
试验地点
3
主要终点
Method comparison: Deming regression analysis (PCM and ROTEM NATEM)

研究概览

简要总结

To determine if PMC viscoelastic coagulation parameters correlate with those generated by a marketed device in adult patients who require hemostasis monitoring or who are at risk of bleeding or thrombotic indications

详细描述

The primary aim of this study is a performance evaluation of the PCM coagulation monitor in comparison to the ROTEM thromboelastometry system.

To include a measurement of the entire range of the test system, subjects with a medical management need that requires the measurement of their hemostasis status (such as PT/PTT, INR) or are at risk for being hypo-coagulable or hyper-coagulable will be included in this study.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age 18 years or older
  • In opinion of investigator, absence of anything that precludes subject from being good study candidate
  • Informed consent
  • Order for measurement of hemostasis status of patients who are at risk for being hyper-coagulable or hypo-coagulable due to clinical diagnosis or medical need

排除标准

  • Previous enrollment in study
  • Incarcerated
  • Heparin infusion in prior 24 hours

结局指标

主要结局

Method comparison: Deming regression analysis (PCM and ROTEM NATEM)

时间窗: Parameter differences assessed at time of blood sampling

Deming regression will be used to compare the PCM and the ROTEM NATEM assay (which reports clotting time, speed of clot formation, maximal clot firmness and clot lysis)

次要结局

  • Correlation between abnormalities(Parameter differences assessed at time of blood sampling)
  • Method comparison: Deming regression analysis (PCM and ROTEM INTEM, EXTEM)(Parameter differences assessed at time of blood sampling)