Perminova Respiratory Effort Accuracy Validation 09-12-2013-Rev 2

Clinimark, LLC1 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2013年9月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
20
试验地点
1
主要终点
Respiratory Rate of test subject as compared to end tidal CO2

研究概览

简要总结

The purpose of this study is to validate the accuracy of the Perminova N4 Monitoring system Respiration Rate technology.

详细描述

After IRB Approval, eighteen volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate technology of the Perminova N4 Monitoring system (Device Under Test). The subjects will be selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.

A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35 and 40 breaths per minute; with some natural variation from these exact numbers.

Each subject will be instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor will be used as the Accuracy Reference Device (Reference) in this study. Performance of the currently released ECG Impedance Respiratory Rate monitor (Acceptance Standard) will be used to establish the acceptance criteria in this study by comparing to Reference.

研究设计

研究类型
Observational
时间视角
Prospective

入排标准

年龄范围
18 Years 至 45 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • range of physiques

排除标准

  • injury to sensor area
  • skin inflammation
  • implanted devices such as pacemakers

结局指标

主要结局

Respiratory Rate of test subject as compared to end tidal CO2

时间窗: 1 week

次要结局

未报告次要终点