临床试验/NCT05673408
NCT05673408招募中不适用

Non-Invasive Blood Pressure Validation Study

Philips Clinical & Medical Affairs Global10 个研究点 分布在 1 个国家目标入组 189 人开始时间: 2023年6月23日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
入组人数
189
试验地点
10
主要终点
Philips NIBP System Equivalence to Radial Arterial Line

研究概览

简要总结

The purpose of this study is to compare measurements of blood pressure (BP) between the Philips non-invasive blood pressure (NIBP) system (including NIBP cuff and portable patient monitor) and invasive radial for Subject Group 2 and 3, and radial or umbilical arterial line for neonates (A-line) in critical care patients.

详细描述

A multi-center, multi-phase, prospective, self-controlled, observational study.

  • Phase I - A pilot phase designed for feasibility and the potential for algorithm development (enhancement).
  • Phase II - Pivotal Phase: At the completion of Phase I -and finalization of the algorithm (if applicable), Phase II is designed for the purpose of establishing equivalence of the Philips NIBP system with the gold standard radial or umbilical (neonates) arterial line.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • Patient admitted into the NICU, PICU, ICU, or OR (operating room)
  • Adult aged 18 years and older, or parent/legal guardian of minor, willing and able to understand and provide informed consent/assent
  • Indication for NIBP cuff
  • Indication for radial arterial line for Subject Group 2 and Subject Group 3, and radial or umbilical arterial line for Subject Group 1
  • Meet the following stratification criteria across the study:
  • Stratification (per the ISO 81060-2:2018/Amd-2:2024):
  • Subject Group 1:
  • At least 3 patients shall be <1000 g in weight
  • At least 3 patients shall be 1000 to 2000 g in weight
  • At least 3 patients shall be > 2000 g in weight
  • At least 3 patients shall be ≥ 29 days and < 1 year of age
  • At least 3 patients shall be ≥ 1 year and < 3 years of age
  • The remaining patients may be from any of the above age or weight groups in order to complete the sample size
  • A patient can be in more than one category simultaneously
  • Subject Group 2:
  • At least 30% male, 30% female
  • Limb circumference distributed as specified in ISO81060-2:2018/Amd2:
  • Subject Group 3:
  • At least 30% male, 30% female
  • Limb circumference distributed as specified in ISO81060- 2:2018/Amd-2:
  • At least 10% SBP ≤ 100mmHg
  • At least 10% SBP ≥ 160mmHg
  • At least 10% DBP ≤ 70mmHg
  • At least 10% DBP ≥ 85mmHg

排除标准

  • Inability to place the study device appropriately due to patient's anatomy or condition
  • Known pregnancy or lactating women (self-report)
  • Patients treated with an intra-aortic balloon pump
  • Aortic and mitral regurgitation (> 2 nd degree)
  • Measurements taken in the lateral position
  • Currently participating in another clinical trial with any product, treatment and/or medication that clinically interferes with the study endpoint(s)
  • If valid SBP reference measurements for lateral difference is > 15 mmHg (except for neonates with umbilical A-line)
  • If valid DBP reference measurements for lateral difference is > 10 mmHg (except for neonates with umbilical A-line)
  • At the Principal Investigator's discretion, subject does not qualify to participate in the study for any reasons that are not mentioned above

结局指标

主要结局

Philips NIBP System Equivalence to Radial Arterial Line

时间窗: Through study completion, approximately 1 year

The primary endpoint is the accuracy and precision of systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP) measured by the Philips NIBP system (including NIBP cuff and portable patient monitor) in comparison to radial arterial line.