Pilot Feasibility and Safety Study on Low-flow Extracorporeal CO2 Removal in Patients With Moderate ARDS to Enhance Lung Protective Ventilation
试验速览
- 阶段
- 1 期
- 状态
- 已完成
- 入组人数
- 95
- 试验地点
- 1
- 主要终点
- Achievement of VT reduction to 4 mL/kg while maintaining pH and PaCO2 to ± 20% of baseline values obtained at VT of 6 mL/kg.
研究概览
简要总结
Pathophysiological, experimental and clinical data suggest that an '"ultraprotective" mechanical ventilation strategy may further reduce VILI and ARDS-associated morbidity and mortality. Severe hypercapnia induced by VT reduction in this setting might be efficiently controlled by ECCO2R devices. A proof-of-concept study conducted on a limited number of ARDS cases indicated that ECCO2R allowing VT reduction to 3.5-5 ml/kg to achieve Pplat<25 cm H2O may further reduce VILI.
详细描述
Over the past few decades, highly significant progress has been made in understanding the pathophysiology of the acute respiratory distress syndrome (ARDS). Recognition of ventilation-induced lung injuries (VILI) has led to the radical modification of the ventilatory management of these patients. The landmark trial by the ARDSnet trial group demonstrated in 2000 that ventilating ARDS patients with a low tidal volume (VT) of 6 ml/kg (calculated from predicted body weight), and with a maximum end-inspiratory plateau pressure (Pplat) of 30 cmH2O decreased mortality from 39.8% (in the conventional arm treated with a VT of 12 ml/kg PBW) to 31% . However, recent studies have shown that lung hyperinflation still occurs in approximately 30% of ARDS patients even though they are being ventilated using the ARDSNet strategy. Additionally, Hager and coworkers found that mortality decreased as Pplat declined from high to low levels at all levels of Pplat on the data collected by the "ARDSNet" trial group. Their analysis suggested a beneficial effect of VT reduction even for patients who already had Pplat<30 cm H2O before VT reduction.Similar observation was also recently reported by Needham et al on a cohort of 485 patients with ARDS. Because VT reduction to <6 ml/kg to achieve very low Pplat may induce severe hypercapnia and may cause elevated intracranial pressure, pulmonary hypertension, decreased myocardial contractility, decreased renal blood flow, and the release of endogenous catecholamines, this strategy using "ultraprotective" MV settings is not possible for most patients on conventional mechanical ventilation for moderate to severe ARDS.
Extracorporeal carbon dioxide removal (ECCO2R) may be used in association with mechanical ventilation to permit VT reduction to <6 ml/kg and to achieve very low Pplat (20-25 cm H2O). In an observational study conducted in the 80's, Gattinoni showed that use of venovenous ECCO2R at a flow of 1.5-2.5 l/min in addition to quasi apneic mechanical ventilation with peak inspiratory pressures limited to 35-45 cmH2O and PEEP set at 15-25 cmH2O resulted in lower than expected mortality in an observational cohort of severe ARDS patients. However, a randomized, controlled single-center study using that same technology and conducted in the 1990s by Morris's group in Utah was stopped early for futility after only 40 patients had been enrolled and failed to demonstrate a mortality benefit with this device (58% in the control group vs. 70% in the treatment group).
In recent years, new-generation ECCO2R devices have been developed. They offer lower resistance to blood flow, have small priming volumes and have much more effective gas exchange. With ECCO2R the patient's PaCO2 is principally determined by the rate of fresh gas flow through the membrane lung. In an ECCO2R animal model, CO2 removal averaged 72±1.2 mL/min at blood flows of 450 mL/min, while CO2 production by the lung decreased by 50% with reduction of minute ventilation from 5.6 L/min at baseline to 2.6 L/min after insertion of the device. Lastly, Terragni et al (15)demonstrated that ECCO2R could improve pulmonary protection by allowing very low tidal volume ventilation (3.5-5 ml/kg of PBW) in a proof-of-concept study of ten patients with ARDS. This strategy was also associated with a significant decrease in pulmonary inflammatory biomarkers.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Mechanical ventilation with expected duration of >24h
- •Moderate ARDS according to the Berlin definition(16) PaO2/FiO2: 200-100 mmHg, with PEEP ≥ 5 cmH2O
排除标准
- •Age <18 years
- •Pregnancy
- •Decompensated heart insufficiency or acute coronary syndrome
- •Severe COPD
- •Major respiratory acidosis PaCO2>60 mmHg
- •Acute brain injury
- •Severe liver insufficiency (Child-Pugh scores >7) or fulminant hepatic failure
- •Heparin-induced thrombocytopenia
- •Contraindication for systemic anticoagulation
- •Patient moribund, decision to limit therapeutic interventions
- •Catheter access to femoral vein or jugular vein impossible
- •Pneumothorax
- •Platelet <50 G/l
研究组 & 干预措施
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: ECCO2R will be initiated during the 2-hour run-in time (Device)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Neuromuscular blocking agents (NMBA) (Other)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Ventilation (Device)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Level of carbon dioxide released at the end of expiration (Other)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Respiratory Rate (Other)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Sweep gas flow (Other)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Ventilation will be adapted (Other)
One single arm
Procedure: Baseline ventilator settings will be established per the EXPRESS protocol: VT = 6 mL/kg (ideal body weight); inspiratory flow will be set at 50-70 L/min resulting in an end-inspiratory pause of 0.2-0.5 sec, I:E ratio 1:1 to 1:3, PEEP set so that the plateau pressure (Pplat), measured during the end-inspiratory pause of 0.2 to 0.5 s, will be within the following limits: 28 cm H2O ≤ Pplat ≤ 30 cm H2O; Set RR to 20-35 to maintain approximately the same minute ventilation as before study initiation. Baseline ventilator settings will be maintained for a 2-hour run-in time (time to setup ECCO2R devices). Use heated humidifiers for gas humidification and minimize instrumental dead space. ECCO2R will be initiated during the 2-hour run-in time. Neuromuscular blocking agents (NMBA) will be used. EtCO2 will be monitored. RR will be kept what it was at Baseline. Sweep gas flow will be adapted. Ventilation will be adapted. Respiratory rate will be adapted.
干预措施: Respiratory rate will be adapted (Other)
结局指标
主要结局
Achievement of VT reduction to 4 mL/kg while maintaining pH and PaCO2 to ± 20% of baseline values obtained at VT of 6 mL/kg.
时间窗: maximum 28 days
次要结局
- Assessment of the changes in pH/ PaO2 /PaCO2(maximum 28 days)
- Amount of CO2 removed by the ECCO2R device(maximum 28 days)
- Evaluation of lung recruitment/derecruitment (FRC measurement by the ventilator, ECHO-LUS…)(maximum 28 days)
- Device CO2 clearance in the first 24 hours of ECCO2R(maximum 28 days)
- The frequency of serious adverse events (SAE).(maximum 28 days)